  • Error Code:[0x0076] ERR LEAK SENS

    «Main Cause» Leak sensor malfunctions (probably due to broken/disconnected wire). «Countermeasure» Turn power OFF and contact a Shimadzu sales or service re... Show Detail

    • No:3579
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x0077] WARNING: LOW SET TEMP

    «Main Cause» For CTO-10AC (vp) or CTO-20AC: If the COOLER MODE is "1" or "2", the room temperature is less than -10ºC. When "0" is set, Settings of oven temp... Show Detail

    • No:3580
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x0079] COOLER NOT EXIST

    «Main Cause» Cooler is not mounted or the cable is snapping. «Countermeasure» Turn power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3581
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x8630] ERR NO RV-* HOME

    «Main Cause» Rotation error in an automatic column-switching valve. «Countermeasure» Turn power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3626
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x8632]

    «Main Cause» Cannot write to oven :Management Device (CMD). «Countermeasure» Once turn off the power, and turn on the power again. If this message still app... Show Detail

    • No:3627
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x8636] ERR POST COOLER

    «Main Cause» The post-column cooler is not rotating correctly. «Countermeasure» If anything is interfering with the fan rotation, turn the power OFF and rem... Show Detail

    • No:3628
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x8637] ERR COOLBLK FAN

    «Main Cause» The block cooling fan is not rotating correctly. «Countermeasure» If anything is interfering with the fan rotation, turn the power OFF and remo... Show Detail

    • No:3629
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x8638] ERR DOOR SENS

    «Main Cause» The door sensor has failed. «Countermeasure» Turn off the power and contact a Shimadzu sales or service representative. Show Detail

    • No:3630
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x8650] ERR OVER T.MAX-L

    «Main Cause» The column oven temperature of heater L has exceeded the value set for [T.MAX]. «Countermeasure» (1) Lower the column oven temperature set valu... Show Detail

    • No:3631
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x0070] ERR OVER T.MAX

    «Main Cause» Temperature control is turned off because the temperature of the column oven exceeded the maximum temperature. «Countermeasure» Settings the in... Show Detail

    • No:3766
    • Date of Issue:2020/12/02 16:32