Frequently Referenced FAQ

FAQ within [ Column Oven ]

11-20 of Total 76

≪ 2 / 8page ≫
  • Error Code:[0x940E] WARNING LOW SET TEMP

    «Main Cause» The setting is too low, relative to room temperature. «Countermeasure» Check the column oven temperature setting in instrument parameter settin... Show Detail

    • No:4047
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x940D] FATAL OVEN TEMP SENSOR L

    «Main Cause» A broken or short circuit may have occurred in the temperature sensor. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu ... Show Detail

    • No:4046
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x940C] FATAL OVEN TEMP SENSOR H

    «Main Cause» A broken or short circuit may have occurred in the temperature sensor. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu ... Show Detail

    • No:4045
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x940B] ERROR OVER T.MAX

    «Main Cause» Temperature control was stopped because the column oven temperature exceeded the upper limit temperature. «Countermeasure» Set the column oven ... Show Detail

    • No:4044
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x9409] FATAL PELTIER UNIT 3

    «Main Cause» The Peltier unit is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service represen... Show Detail

    • No:4043
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x9408] FATAL INSIDE FAN

    «Main Cause» The fan inside the compartment may have stopped. «Countermeasure» If tubing or other foreign matter is obstructing fan rotation, switch the ins... Show Detail

    • No:4042
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x9407] FATAL PELTIER FAN

    «Main Cause» A rotation problem was detected for the Peltier unit fan (outside the oven cabinet). «Countermeasure» Switch the instrument power OFF and conta... Show Detail

    • No:4041
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x9406] FATAL PELTIER UNIT 2

    «Main Cause» A broken circuit may have occurred in the Peltier unit. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or servic... Show Detail

    • No:4040
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x9405] FATAL PELTIER UNIT 1

    «Main Cause» A failure occurred in the Peltier unit control circuit. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or servic... Show Detail

    • No:4039
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x9404] FATAL THERMAL FUSE

    «Main Cause» A broken circuit may have occurred in the thermal fuse. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or servic... Show Detail

    • No:4038
    • Date of Issue:2020/12/02 16:45