• Font Size
  • S
  • M
  • L

FAQ within [ Column Oven ]

11-20 of Total 76

2 / 8page
  • Error Code:[0x5629] ERROR DOOR SENSOR

    «Main Cause» The oven door was left ajar for a long time. «Countermeasure» Make sure the door is securely closed. If it is fully closed, it might be a door ... Show Detail

    • No:3395
    • Date of Issue:2020/12/02 15:47
  • Error Code:[0x562A] FATAL AC INPUT POWER

    «Main Cause» The power supply voltage or frequency was not recognized properly. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sale... Show Detail

    • No:3396
    • Date of Issue:2020/12/02 15:47
  • Error Code:[0x562B] ERROR OVER T.MAX

    «Main Cause» Temperature control was stopped because the oven temperature exceeded the upper limit temperature. «Countermeasure» Set the oven temperature se... Show Detail

    • No:3397
    • Date of Issue:2020/12/02 15:47
  • Error Code:[0x562C] ERROR LEAK DETECTED

    «Main Cause» Temperature control was stopped because leakage was detected in the oven. «Countermeasure» Inspect the leakage from flow channels or other loca... Show Detail

    • No:3398
    • Date of Issue:2020/12/02 15:47
  • Error Code:[0x562D] ERROR GAS DETECTED

    «Main Cause» Temperature control was stopped because the gas sensor inside the oven detected organic solvent leakage. «Countermeasure» (1) If an organic sol... Show Detail

    • No:3399
    • Date of Issue:2020/12/02 15:47
  • Error Code:[0x562E] ERROR EXTERNAL INPUT

    «Main Cause» Temperature control stopped because an error signal was input via the oven external input terminal. «Countermeasure» Check the connected extern... Show Detail

    • No:3400
    • Date of Issue:2020/12/02 15:47
  • Error Code:[0x562F] FATAL LEAK SENSOR

    «Main Cause» The leak sensor cannot be detected properly. A broken or short circuit may have occurred. «Countermeasure» Switch the instrument power OFF and ... Show Detail

    • No:3401
    • Date of Issue:2020/12/02 15:47
  • Error Code:[0x5630] WARNING LOW SET TEMP

    «Main Cause» The setting is too low, relative to room temperature. «Countermeasure» Check the oven temperature setting in instrument parameter settings. If ... Show Detail

    • No:3402
    • Date of Issue:2020/12/02 15:47
  • Error Code:[0x9217]

    «Main Cause» The backup parameters for the autosampler and column oven are corrupted. «Countermeasure» Press the (power switch) to turn the power OFF and co... Show Detail

    • No:3485
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x9218]

    «Main Cause» The consumable part count data for the autosampler and column oven is corrupted. «Countermeasure» Press the (power switch) to turn the power OF... Show Detail

    • No:3486
    • Date of Issue:2020/12/02 16:10

11-20 of Total 76

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.