  • (LC) Gas Sensor Calibration

    If an error message fails to appear when the mobile phase leaks, or if an error message appears, althoughthere is no leakage, the leak sensor sensitivity mus... Show Detail

    • No:8576
    • Date of Issue:2020/10/30 11:19
    • Brached FAQ
  • Error Code:[0x5620] FATAL ROM FAILURE

    «Main Cause» A system error occurred. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3386
    • Date of Issue:2020/12/02 15:47
  • Error Code:[0x5621] FATAL RAM FAILURE

    «Main Cause» A system error occurred. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3387
    • Date of Issue:2020/12/02 15:47
  • Error Code:[0x5622] FATAL THERMAL FUSE

    «Main Cause» A broken circuit may have occurred in the thermal fuse. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or servic... Show Detail

    • No:3388
    • Date of Issue:2020/12/02 15:47
  • Error Code:[0x5623] FATAL INSIDE FAN

    «Main Cause» The fan inside the compartment may have stopped. «Countermeasure» If tubing or other foreign matter is obstructing fan rotation, switch the ins... Show Detail

    • No:3389
    • Date of Issue:2020/12/02 15:47
  • Error Code:[0x5624] FATAL OVEN TEMP SENSOR H

    «Main Cause» A broken or short circuit may have occurred in the temperature sensor. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu ... Show Detail

    • No:3390
    • Date of Issue:2020/12/02 15:47
  • Error Code:[0x5625] FATAL OVEN TEMP SENSOR L

    «Main Cause» A broken or short circuit may have occurred in the temperature sensor. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu ... Show Detail

    • No:3391
    • Date of Issue:2020/12/02 15:47
  • Error Code:[0x5626] FATAL GAS SENSOR

    «Main Cause» A broken circuit may have occurred in the gas sensor. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service ... Show Detail

    • No:3392
    • Date of Issue:2020/12/02 15:47
  • Error Code:[0x5627] FATAL AC HEATER

    «Main Cause» The heater is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3393
    • Date of Issue:2020/12/02 15:47
  • Error Code:[0x5628] FATAL ROOM TEMP SENSOR

    «Main Cause» A broken or short circuit may have occurred in the room temperature sensor. «Countermeasure» Switch the instrument power OFF and contact a Shim... Show Detail

    • No:3394
    • Date of Issue:2020/12/02 15:47