FAQ - Frequently Asked Questions
- Products Search > HPLC > Autosampler
Keyword Search
Frequently Referenced FAQ
FAQ within [ Autosampler ]
-
Error Code:[0x8542] NOT ADJUSTED-M
«Main Cause» A sample rack of which rack position has not been adjusted is inserted in rack plate M. «Countermeasure» Adjust the rack position. Show Detail
- No:3593
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x8543] NOT ADJUSTED-R
«Main Cause» A sample rack of which rack position has not been adjusted is inserted in rack plate R. «Countermeasure» Adjust the rack position. Show Detail
- No:3594
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x8544] NO RACK ID
«Main Cause» Rack ID is not set correctly. «Countermeasure» Set a rack ID correctly. Show Detail
- No:3595
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x8546] STOP COOLER L
«Main Cause» Since sample cooler L has operated without sample rack for more than 30 minutes, sample cooler L is stopped. «Countermeasure» Insert a sample r... Show Detail
- No:3596
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x8547] STOP COOLER M
«Main Cause» Since sample cooler M has operated without sample rack for more than 30 minutes, sample cooler M is stopped. «Countermeasure» Insert a sample r... Show Detail
- No:3597
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x8548] STOP COOLER R
«Main Cause» Since sample cooler R has operated without sample rack for more than 30 minutes, sample cooler R is stopped. «Countermeasure» Insert a sample r... Show Detail
- No:3598
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x8550] ERR LEAK DETECT
«Main Cause» (1) Leak sensor was detected leakage. (2) Sensor is too sensitive. «Countermeasure» (1) Inspect the leakage from the flow line. Wipe out spille... Show Detail
- No:3599
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x8551] ERR LEAK DETECT
«Main Cause» (1) Leak sensor was detected leakage. (2) Sensor is too sensitive. «Countermeasure» (1) Inspect the leakage from the flow line. Wipe out spille... Show Detail
- No:3600
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x8563] ERR SLIP X
«Main Cause» Needle's X-axis (sideways) movement is incorrect. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:3601
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x8564] ERR SLIP Y
«Main Cause» Needle's Y-axis (forward/backward) movement is incorrect. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service represen... Show Detail
- No:3602
- Date of Issue:2020/12/02 16:30
81-90 of Total 297