• Font Size
  • S
  • M
  • L

FAQ within [ Autosampler ]

81-90 of Total 297

9 / 30page
  • Error Code:[0x8542] NOT ADJUSTED-M

    «Main Cause» A sample rack of which rack position has not been adjusted is inserted in rack plate M. «Countermeasure» Adjust the rack position. Show Detail

    • No:3593
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x8543] NOT ADJUSTED-R

    «Main Cause» A sample rack of which rack position has not been adjusted is inserted in rack plate R. «Countermeasure» Adjust the rack position. Show Detail

    • No:3594
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x8544] NO RACK ID

    «Main Cause» Rack ID is not set correctly. «Countermeasure» Set a rack ID correctly. Show Detail

    • No:3595
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x8546] STOP COOLER L

    «Main Cause» Since sample cooler L has operated without sample rack for more than 30 minutes, sample cooler L is stopped. «Countermeasure» Insert a sample r... Show Detail

    • No:3596
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x8547] STOP COOLER M

    «Main Cause» Since sample cooler M has operated without sample rack for more than 30 minutes, sample cooler M is stopped. «Countermeasure» Insert a sample r... Show Detail

    • No:3597
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x8548] STOP COOLER R

    «Main Cause» Since sample cooler R has operated without sample rack for more than 30 minutes, sample cooler R is stopped. «Countermeasure» Insert a sample r... Show Detail

    • No:3598
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x8550] ERR LEAK DETECT

    «Main Cause» (1) Leak sensor was detected leakage. (2) Sensor is too sensitive. «Countermeasure» (1) Inspect the leakage from the flow line. Wipe out spille... Show Detail

    • No:3599
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x8551] ERR LEAK DETECT

    «Main Cause» (1) Leak sensor was detected leakage. (2) Sensor is too sensitive. «Countermeasure» (1) Inspect the leakage from the flow line. Wipe out spille... Show Detail

    • No:3600
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x8563] ERR SLIP X

    «Main Cause» Needle's X-axis (sideways) movement is incorrect. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3601
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x8564] ERR SLIP Y

    «Main Cause» Needle's Y-axis (forward/backward) movement is incorrect. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service represen... Show Detail

    • No:3602
    • Date of Issue:2020/12/02 16:30

81-90 of Total 297

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.