Error Code:[0x0015] SIL:NO SYRINGE HOME
«Main Cause» Motion of syringe plunger driving motor was faulty. «Countermeasure» The friction between the syringe tip and barrel may have excessively incr... Show Detail
- No:3562
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x0016] SIL:PFILE PRG ERROR
«Main Cause» There is a line which cannot be executed in the pretreatment program. «Countermeasure» Check set parameter for the line. Show Detail
- No:3563
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x0017] SIL:PFILE NOT EXIST
«Main Cause» Pretreatment program to be executed does not exist. «Countermeasure» Check if file number to be executed is specified correctly. Show Detail
- No:3564
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x001A] SIL:NO 6-PORT VALVE HOME
«Main Cause» 6-port valve does not rotate to the correct positions. «Countermeasure» Please perform actions using the following procedures. (1) Check that ... Show Detail
- No:3565
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x001B] ERR PUMP HOME
«Main Cause» Metering pump home position error has occurred. «Countermeasure» Remedy this problem referring to the following procedure. Check if the plunge... Show Detail
- No:3566
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x001C] SIL:SYRINGE VOLUME
«Main Cause» Suction of more volume than syringe capacity was attempted in pretreatment program. «Countermeasure» If suction volume is larger than syringe ... Show Detail
- No:3567
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x001E] SIL:SYRINGE VOLUME
«Main Cause» Invalid aspiration/discharge volume setting of metering pump in pretreatment program. «Countermeasure» If suction volume is larger than syring... Show Detail
- No:3568
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x0091] NO VIAL DETECTED
«Main Cause» No sample vial was placed in the rack position. «Countermeasure» Remedy this problem referring to the following procedure. (1) The sample via... Show Detail
- No:3569
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x0092] ERR LEAK DETECT
«Main Cause» Concentration of organic phase vapor inside the unit has exceeded the leak sensor actuation level. «Countermeasure» Check for liquid leakage f... Show Detail
- No:3570
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x8513] ERR LPV HOME
«Main Cause» Low-pressure valve's rotational position incorrect positions. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service repr... Show Detail
- No:3582
- Date of Issue:2020/12/02 16:30