• Font Size
  • S
  • M
  • L

Frequently Referenced FAQ

FAQ within [ Autosampler ]

61-70 of Total 299

≪ 7 / 30page ≫
  • Error Code:[0x0015] SIL:NO SYRINGE HOME

    «Main Cause» Motion of syringe plunger driving motor was faulty. «Countermeasure» The friction between the syringe tip and barrel may have excessively incr... Show Detail

    • No:3562
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x0016] SIL:PFILE PRG ERROR

    «Main Cause» There is a line which cannot be executed in the pretreatment program. «Countermeasure» Check set parameter for the line. Show Detail

    • No:3563
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x0017] SIL:PFILE NOT EXIST

    «Main Cause» Pretreatment program to be executed does not exist. «Countermeasure» Check if file number to be executed is specified correctly. Show Detail

    • No:3564
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x001A] SIL:NO 6-PORT VALVE HOME

    «Main Cause» 6-port valve does not rotate to the correct positions. «Countermeasure» Please perform actions using the following procedures. (1) Check that ... Show Detail

    • No:3565
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x001B] ERR PUMP HOME

    «Main Cause» Metering pump home position error has occurred. «Countermeasure» Remedy this problem referring to the following procedure. Check if the plunge... Show Detail

    • No:3566
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x001C] SIL:SYRINGE VOLUME

    «Main Cause» Suction of more volume than syringe capacity was attempted in pretreatment program. «Countermeasure» If suction volume is larger than syringe ... Show Detail

    • No:3567
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x001E] SIL:SYRINGE VOLUME

    «Main Cause» Invalid aspiration/discharge volume setting of metering pump in pretreatment program. «Countermeasure» If suction volume is larger than syring... Show Detail

    • No:3568
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x0091] NO VIAL DETECTED

    «Main Cause» No sample vial was placed in the rack position. «Countermeasure» Remedy this problem referring to the following procedure. (1) The sample via... Show Detail

    • No:3569
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x0092] ERR LEAK DETECT

    «Main Cause» Concentration of organic phase vapor inside the unit has exceeded the leak sensor actuation level. «Countermeasure» Check for liquid leakage f... Show Detail

    • No:3570
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x8513] ERR LPV HOME

    «Main Cause» Low-pressure valve's rotational position incorrect positions. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service repr... Show Detail

    • No:3582
    • Date of Issue:2020/12/02 16:30

61-70 of Total 299

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Technology Inc.