Error Code:[0x527F] FATAL PCG FAN INNER
«Main Cause» A fan rotation problem was detected for the fan inside the cabinet. «Countermeasure» Check for foreign matter caught in the fan in the cabinet.... Show Detail
- No:4008
- Date of Issue:2020/12/02 16:45
-
Error Code:[0x527E] ERROR PCG DOOR OPEN
«Main Cause» Temperature control was stopped because the front door was open for a given period. «Countermeasure» Close the front door and start temperature... Show Detail
- No:4007
- Date of Issue:2020/12/02 16:45
-
Error Code:[0x527D] WARNING PCG DOOR OPEN
«Main Cause» The front door is open. «Countermeasure» Close the front door. Show Detail
- No:4006
- Date of Issue:2020/12/02 16:45
-
Error Code:[0x527C] FATAL PCG PELTIER2 VOL
«Main Cause» The Peltier element used for temperature adjustment is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a... Show Detail
- No:4005
- Date of Issue:2020/12/02 16:45
-
Error Code:[0x527B] FATAL PCG PELTIER1 VOL
«Main Cause» The Peltier element used for temperature adjustment is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a... Show Detail
- No:4004
- Date of Issue:2020/12/02 16:45
-
Error Code:[0x527A] ERROR PCG TEMP SENSOR2
«Main Cause» Temperature sensor 2 inside the cabinet detected a temperature outside the specified range. The temperature sensor malfunctioned or the temperat... Show Detail
- No:4003
- Date of Issue:2020/12/02 16:45
-
Error Code:[0x5279] ERROR PCG TEMP SENSOR1
«Main Cause» Temperature sensor 1 inside the cabinet detected a temperature outside the specified range. The temperature sensor malfunctioned or the temperat... Show Detail
- No:4002
- Date of Issue:2020/12/02 16:45
-
Error Code:[0x5278] ERROR PCG PLATE FALL
«Main Cause» A sample plate dropped during transfer. «Countermeasure» Open the front door and recover the dropped sample plate. Show Detail
- No:4001
- Date of Issue:2020/12/02 16:45
-
Error Code:[0x5277] ERROR PCG PLATE CHUCK
«Main Cause» The arm chuck cannot grip the sample plate. «Countermeasure» Make sure the sample plate is placed correctly. Show Detail
- No:4000
- Date of Issue:2020/12/02 16:45
-
Error Code:[0x5276] ERROR PCG NO PLATE
«Main Cause» The specified sample plate was not found. «Countermeasure» Place the sample plate in the specified stacker. Show Detail
- No:3999
- Date of Issue:2020/12/02 16:45