FAQ - Frequently Asked Questions
- Products Search > HPLC > Autosampler
Keyword Search
Frequently Referenced FAQ
FAQ within [ Autosampler ]
-
Error Code:[0x527E] ERROR PCG DOOR OPEN
«Main Cause» Temperature control was stopped because the front door was open for a given period. «Countermeasure» Close the front door and start temperature... Show Detail
- No:4007
- Date of Issue:2020/12/02 16:45
-
Error Code:[0x527D] WARNING PCG DOOR OPEN
«Main Cause» The front door is open. «Countermeasure» Close the front door. Show Detail
- No:4006
- Date of Issue:2020/12/02 16:45
-
Error Code:[0x527C] FATAL PCG PELTIER2 VOL
«Main Cause» The Peltier element used for temperature adjustment is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a... Show Detail
- No:4005
- Date of Issue:2020/12/02 16:45
-
Error Code:[0x527B] FATAL PCG PELTIER1 VOL
«Main Cause» The Peltier element used for temperature adjustment is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a... Show Detail
- No:4004
- Date of Issue:2020/12/02 16:45
-
Error Code:[0x527A] ERROR PCG TEMP SENSOR2
«Main Cause» Temperature sensor 2 inside the cabinet detected a temperature outside the specified range. The temperature sensor malfunctioned or the temperat... Show Detail
- No:4003
- Date of Issue:2020/12/02 16:45
-
Error Code:[0x5279] ERROR PCG TEMP SENSOR1
«Main Cause» Temperature sensor 1 inside the cabinet detected a temperature outside the specified range. The temperature sensor malfunctioned or the temperat... Show Detail
- No:4002
- Date of Issue:2020/12/02 16:45
-
Error Code:[0x5278] ERROR PCG PLATE FALL
«Main Cause» A sample plate dropped during transfer. «Countermeasure» Open the front door and recover the dropped sample plate. Show Detail
- No:4001
- Date of Issue:2020/12/02 16:45
-
Error Code:[0x5277] ERROR PCG PLATE CHUCK
«Main Cause» The arm chuck cannot grip the sample plate. «Countermeasure» Make sure the sample plate is placed correctly. Show Detail
- No:4000
- Date of Issue:2020/12/02 16:45
-
Error Code:[0x5276] ERROR PCG NO PLATE
«Main Cause» The specified sample plate was not found. «Countermeasure» Place the sample plate in the specified stacker. Show Detail
- No:3999
- Date of Issue:2020/12/02 16:45
-
Error Code:[0x5275] ERROR PCG PLATE FULL
«Main Cause» The sample plate cannot be returned because there are no empty positions in the stack. «Countermeasure» Pull out the sample plate from the stack. Show Detail
- No:3998
- Date of Issue:2020/12/02 16:45
51-60 of Total 297