• Font Size
  • S
  • M
  • L

Frequently Referenced FAQ

FAQ within [ Autosampler ]

41-50 of Total 299

≪ 5 / 30page ≫
  • Error Code:[0x920D]

    «Main Cause» The leak sensor in the autosampler high pressure valve has detected an abnormal condition. «Countermeasure» Press the (power switch) to turn th... Show Detail

    • No:3477
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x920E]

    «Main Cause» The leak sensor in the autosampler high pressure valve is malfunctioning. «Countermeasure» Press the (power switch) to turn the power OFF and c... Show Detail

    • No:3478
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x920F]

    «Main Cause» Temperature control is turned off because the temperature sensor in autosampler rack L is malfunctioning. «Countermeasure» Press the (power swi... Show Detail

    • No:3479
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x9210]

    «Main Cause» Temperature control is turned off because the temperature sensor in autosampler rack L has detected an abnormal condition. «Countermeasure» Pre... Show Detail

    • No:3480
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x9213]

    «Main Cause» Temperature control is turned off because the temperature sensor in autosampler rack R is malfunctioning. «Countermeasure» Press the (power swi... Show Detail

    • No:3481
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x9214]

    «Main Cause» Temperature control is turned off because the temperature sensor in autosampler rack R has detected an abnormal condition. «Countermeasure» Pre... Show Detail

    • No:3482
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x9215]

    «Main Cause» Temperature control is turned off because the temperature sensor in the autosampler dehumidifier unit is malfunctioning. «Countermeasure» Press... Show Detail

    • No:3483
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x9216]

    «Main Cause» Temperature control is turned off because the sensor in the autosampler dehumidifier unit has detected an abnormal condition. «Countermeasure» ... Show Detail

    • No:3484
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x9217]

    «Main Cause» The backup parameters for the autosampler and column oven are corrupted. «Countermeasure» Press the (power switch) to turn the power OFF and co... Show Detail

    • No:3485
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x9218]

    «Main Cause» The consumable part count data for the autosampler and column oven is corrupted. «Countermeasure» Press the (power switch) to turn the power OF... Show Detail

    • No:3486
    • Date of Issue:2020/12/02 16:10

41-50 of Total 299

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Technology Inc.