Frequently Referenced FAQ

FAQ within [ Autosampler ]

31-40 of Total 299

4 / 30page
  • Error Code:[0x5293] ERROR FCVx IS LOCAL

    «Main Cause» FCV became local mode during analysis. «Countermeasure» Check if the remote LED of the FCV is lit. Show Detail

    • No:4153
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x5292] ERROR FCVx DISCONNECED

    «Main Cause» The cause is that communication with the FCV whose [FCV FIX] setting is [1: ON] is cut off. «Countermeasure» Make sure the FCV is turned on. Ch... Show Detail

    • No:4152
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x5291] ERROR ACQUISITION TIME

    «Main Cause» Autosampler: Repeated injections did not complete within the data acquisition time. «Countermeasure» Set the injection interval to a time longe... Show Detail

    • No:4151
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x5290] FATAL SYRINGE DATA

    «Main Cause» Autosampler: The data stored in the syringe unit is not correct, so the data was initialized. «Countermeasure» Although this does not affect th... Show Detail

    • No:4150
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x528F] ERROR SYRINGE LEAK

    «Main Cause» Autosampler: The syringe unit detected leak. «Countermeasure» Wipe off the leaking liquid and check the leaking part. If fluid is leaking from ... Show Detail

    • No:4149
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x528E] FATAL SYRINGE LEAK SEN

    «Main Cause» Autosampler: There is an error in the leak sensor of the syringe unit. The leakage sensor may be broken. «Countermeasure» Turn off the instrume... Show Detail

    • No:4148
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x528D] FATAL SYRINGE LPV HOME

    «Main Cause» Autosampler: The 3 way valve of the syringe unit failed to detect the home position. «Countermeasure» Turn off the instrument and contact a Shi... Show Detail

    • No:4147
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x528C] FATAL SYRINGE HOME

    «Main Cause» Autosampler: The syringe unit failed to detect the home position. «Countermeasure» Check the syringe plunger and plunger holder for foreign obj... Show Detail

    • No:4146
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x81E5] FATAL LEAK SENSOR

    «Main Cause» The leak sensor cannot be detected properly. The leak sensor might not be connected. «Countermeasure» Switch the instrument power OFF and conta... Show Detail

    • No:4021
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x528B] WARNING SMP, L: -

    «Main Cause» An error occurred in the sample table being executed. The row number of the row with an error is indicated to the right of "L:" and the error ty... Show Detail

    • No:4020
    • Date of Issue:2020/12/02 16:45