• Font Size
  • S
  • M
  • L

Frequently Referenced FAQ

FAQ within [ Autosampler ]

21-30 of Total 299

≪ 3 / 30page ≫
  • Error Code:[0x922C]

    «Main Cause» The rack ID is not correct. «Countermeasure» Set the correct rack ID for the rack. Show Detail

    • No:3457
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x922D]

    «Main Cause» Racks that have the same ID are set in the autosampler. «Countermeasure» Change the rack ID of either the left or right side rack. Show Detail

    • No:3458
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x922E]

    «Main Cause» The left side sample cooler temperature adjustment was stopped because the left side rack was not set for more than 30 minutes. «Count... Show Detail

    • No:3459
    • Date of Issue:2020/12/02 16:10
    • Last Updated:2022/01/06 09:43
  • Error Code:[0x922F]

    «Main Cause» The left side sample cooler temperature adjustment was stopped because the right side rack was not set for more than 30 minutes. «Countermeasur... Show Detail

    • No:3460
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x9230]

    «Main Cause» The sample vial number that is outside of the setting range for the set plate is specified in the sequence. «Countermeasure» Check the sequence... Show Detail

    • No:3461
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x9231]

    «Main Cause» The autosampler needle X-direction motor has slipped. «Countermeasure» Press the (power switch) to turn the power OFF and contact a Shimadzu sa... Show Detail

    • No:3462
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x9232]

    «Main Cause» The autosampler needle Y-direction motor has slipped. «Countermeasure» Press the (power switch) to turn the power OFF and contact a Shimadzu sa... Show Detail

    • No:3463
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x9233]

    «Main Cause» The remaining autosampler rinse solution is low. «Countermeasure» Replenish the autosampler rinse solution. Show Detail

    • No:3464
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x9234]

    «Main Cause» When the autosampler needle was lowered, it hit a foreign object causing excessive force to be exerted on it. «Countermeasure» Press the (power... Show Detail

    • No:3465
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x9200]

    «Main Cause» There is a mistake in the analysis parameters or settings. «Countermeasure» Check whether there are any mistakes or discrepancies in the acquis... Show Detail

    • No:3466
    • Date of Issue:2020/12/02 16:10

21-30 of Total 299

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Technology Inc.