FAQ within [ Autosampler ]

281-290 of Total 297

29 / 30page
  • Error Code:[0x9228]

    «Main Cause» Temperature control is turned off because the temperature sensor in autosampler rack plate R has detected an abnormal condition. «Countermeasur... Show Detail

    • No:3453
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x9227]

    «Main Cause» Temperature control is turned off because the temperature sensor in autosampler rack plate L has detected an abnormal condition. «Countermeasur... Show Detail

    • No:3452
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x9225]

    «Main Cause» The temperature sensor in the autosampler dehumidifier unit has detected an abnormal value. «Countermeasure» Press the (power switch) to turn t... Show Detail

    • No:3451
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x9224]

    «Main Cause» System Error «Countermeasure» Press the (power switch) to turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3450
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x9223]

    «Main Cause» Could not detect the home position of the autosampler low pressure valve. «Countermeasure» Press the (power switch) to turn the power OFF and c... Show Detail

    • No:3449
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x9222]

    «Main Cause» There is no sample vial set at the sample vial number specified in the sequence. «Countermeasure» Set a sample vial at the sample vial number s... Show Detail

    • No:3448
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x9221]

    «Main Cause» Could not detect the forward and backward home position sensor of the autosampler measuring pump. «Countermeasure» Press the (power switch) to ... Show Detail

    • No:3447
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x9220]

    «Main Cause» Could not detect the Z-direction home position of the autosampler needle. «Countermeasure» Press the (power switch) to turn the power OFF and c... Show Detail

    • No:3446
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x921F]

    «Main Cause» Could not detect the Y-direction home position of the autosampler needle. «Countermeasure» Press the (power switch) to turn the power OFF and c... Show Detail

    • No:3445
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x921E]

    «Main Cause» Could not detect the X-direction home position of the autosampler needle. «Countermeasure» Press the (power switch) to turn the power OFF and c... Show Detail

    • No:3444
    • Date of Issue:2020/12/02 16:10