FAQ - Frequently Asked Questions
- Products Search > HPLC > Autosampler
Keyword Search
Frequently Referenced FAQ
FAQ within [ Autosampler ]
-
«Main Cause» Temperature control is turned off because the autosampler rack L heater has overheated and the thermal fuse has blown. «Countermeasure» Press t... Show Detail
- No:3473
- Date of Issue:2020/12/02 16:10
-
«Main Cause» Temperature control is turned off because voltage of the autosampler dehumidifier unit is too high. «Countermeasure» Press the (power switch) t... Show Detail
- No:3472
- Date of Issue:2020/12/02 16:10
-
«Main Cause» Temperature control is turned off because voltage of the autosampler rack R cooling unit is too high. «Countermeasure» Press the (power switch)... Show Detail
- No:3471
- Date of Issue:2020/12/02 16:10
-
«Main Cause» Temperature control is turned off because voltage of the autosampler rack L cooling unit is too high. «Countermeasure» Press the (power switch)... Show Detail
- No:3470
- Date of Issue:2020/12/02 16:10
-
«Main Cause» Could not detect the rotation direction home position of the autosampler measuring pump. «Countermeasure» Press the (power switch) to turn the ... Show Detail
- No:3469
- Date of Issue:2020/12/02 16:10
-
«Main Cause» There is a liquid leak from the high pressure valve, or from a unit positioned above the high pressure value (column oven or detector, etc.). «... Show Detail
- No:3468
- Date of Issue:2020/12/02 16:10
-
«Main Cause» There is a liquid leak from the low pressure valve or metering pump, or from a unit positioned above the low pressure value (high pressure valve... Show Detail
- No:3467
- Date of Issue:2020/12/02 16:10
-
«Main Cause» There is a mistake in the analysis parameters or settings. «Countermeasure» Check whether there are any mistakes or discrepancies in the acquis... Show Detail
- No:3466
- Date of Issue:2020/12/02 16:10
-
«Main Cause» When the autosampler needle was lowered, it hit a foreign object causing excessive force to be exerted on it. «Countermeasure» Press the (power... Show Detail
- No:3465
- Date of Issue:2020/12/02 16:10
-
«Main Cause» The remaining autosampler rinse solution is low. «Countermeasure» Replenish the autosampler rinse solution. Show Detail
- No:3464
- Date of Issue:2020/12/02 16:10
261-270 of Total 297