• Font Size
  • S
  • M
  • L

FAQ within [ Autosampler ]

241-250 of Total 297

25 / 30page
  • Error Code:[0x0011] ERR NDLE HOME Y

    «Main Cause» Move to Y (front and back) direction was faulty. «Countermeasure» Check if nothing is stuck in the movable part. If nothing is stuck, contact ... Show Detail

    • No:3558
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x0010] ERR NDLE HOME X

    «Main Cause» Move to X (right and left) direction was faulty. «Countermeasure» Check if nothing is stuck in the movable part. If nothing is stuck, contact ... Show Detail

    • No:3557
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x921D]

    «Main Cause» Teaching has not been executed for the autosampler injection port. «Countermeasure» Adjust the position of the needle in relation t... Show Detail

    • No:3491
    • Date of Issue:2020/12/02 16:10
    • Last Updated:2021/01/19 12:52
  • Error Code:[0x921C]

    «Main Cause» Could not detect the home position of the autosampler high pressure valve. «Countermeasure» Press the (power switch) to turn the power OFF and ... Show Detail

    • No:3490
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x921B]

    «Main Cause» A rotation error was detected in the exhaust heat fan at the top of the rear of the instrument. «Countermeasure» Press the (power switch) to tu... Show Detail

    • No:3489
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x921A]

    «Main Cause» Temperature control is turned off because a rotation error was detected in the cooling fan of the autosampler. «Countermeasure» Press the (powe... Show Detail

    • No:3488
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x9219]

    «Main Cause» The front panel of the autosampler may not be attached correctly. Alternatively, the sensor that detects the position of the front panel is malf... Show Detail

    • No:3487
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x9218]

    «Main Cause» The consumable part count data for the autosampler and column oven is corrupted. «Countermeasure» Press the (power switch) to turn the power OF... Show Detail

    • No:3486
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x9217]

    «Main Cause» The backup parameters for the autosampler and column oven are corrupted. «Countermeasure» Press the (power switch) to turn the power OFF and co... Show Detail

    • No:3485
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x9216]

    «Main Cause» Temperature control is turned off because the sensor in the autosampler dehumidifier unit has detected an abnormal condition. «Countermeasure» ... Show Detail

    • No:3484
    • Date of Issue:2020/12/02 16:10

241-250 of Total 297

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.