Error Code:[0x526E] ERROR PCG Y HOME
«Main Cause» Something is obstructing the arm's line of movement or there is mechanical problem in the drive unit. «Countermeasure» Check for any obstructio... Show Detail
- No:3991
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x526F] ERROR PCG Y SLIP
«Main Cause» Something is obstructing the arm's line of movement or there is mechanical problem in the drive unit. «Countermeasure» Check for any obstructio... Show Detail
- No:3992
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x5270] ERROR PCG CHUCK HOME
«Main Cause» Something is obstructing the arm's line of movement or there is mechanical problem in the drive unit. «Countermeasure» Check for any obstructio... Show Detail
- No:3993
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x5271] ERROR PCG CHUCK OPEN
«Main Cause» Something is obstructing the arm's line of movement or there is mechanical problem in the drive unit. «Countermeasure» Check for any obstructio... Show Detail
- No:3994
- Date of Issue:2020/12/02 16:45
-
Error Code:[0x5272] ERROR PCG CHUCK CLOSE
«Main Cause» Something is obstructing the arm's line of movement or there is mechanical problem in the drive unit. «Countermeasure» Check for any obstructio... Show Detail
- No:3995
- Date of Issue:2020/12/02 16:45
-
Error Code:[0x5273] ERROR PCG PLATE BACK
«Main Cause» The arm chuck cannot release the sample plate. «Countermeasure» Restart the instrument. Show Detail
- No:3996
- Date of Issue:2020/12/02 16:45
-
Error Code:[0x5274] ERROR PCG PLATE SET
«Main Cause» A separate sample plate was placed in the stack where the plate in the autosampler was previously placed. «Countermeasure» Check the sample plate. Show Detail
- No:3997
- Date of Issue:2020/12/02 16:45
-
Error Code:[0x5275] ERROR PCG PLATE FULL
«Main Cause» The sample plate cannot be returned because there are no empty positions in the stack. «Countermeasure» Pull out the sample plate from the stack. Show Detail
- No:3998
- Date of Issue:2020/12/02 16:45
-
Error Code:[0x5276] ERROR PCG NO PLATE
«Main Cause» The specified sample plate was not found. «Countermeasure» Place the sample plate in the specified stacker. Show Detail
- No:3999
- Date of Issue:2020/12/02 16:45
-
Error Code:[0x5277] ERROR PCG PLATE CHUCK
«Main Cause» The arm chuck cannot grip the sample plate. «Countermeasure» Make sure the sample plate is placed correctly. Show Detail
- No:4000
- Date of Issue:2020/12/02 16:45