      Error Code:[0x8540] NOT ADJUST RACK «Main Cause» Analysis is started on the sample rack of which rack position has not been adjusted. «Countermeasure» Adjust the rack position. Show Detail - No:3591
- Date of Issue:2020/12/02 16:30
 
- 
      Error Code:[0x8539] ERR TEMP SENS R «Main Cause» Sample cooler R temperature sensor error. «Countermeasure» Turn off the power and contact a Shimadzu sales or service representative. Show Detail - No:3590
- Date of Issue:2020/12/02 16:30
 
- 
      Error Code:[0x8538] ERR TEMP SENS M «Main Cause» Sample cooler M temperature sensor error. «Countermeasure» Turn off the power and contact a Shimadzu sales or service representative. Show Detail - No:3589
- Date of Issue:2020/12/02 16:30
 
- 
      Error Code:[0x8537] ERR TEMP SENS L «Main Cause» Sample cooler L temperature sensor error. «Countermeasure» Turn off the power and contact a Shimadzu sales or service representative. Show Detail - No:3588
- Date of Issue:2020/12/02 16:30
 
- 
      Error Code:[0x8535] ERR TEMP SENS RM «Main Cause» (1) The room temperature sensor detects an abnormally high or low temperature. (2) Room temperature sensor error. «Countermeasure» (1) If the r... Show Detail - No:3587
- Date of Issue:2020/12/02 16:30
 
- 
      Error Code:[0x8534] ERR TEMP SENS DH «Main Cause» Peltier dehumidifier temperature sensor error. «Countermeasure» Turn power OFF and contact a Shimadzu sales or service representative. Show Detail - No:3586
- Date of Issue:2020/12/02 16:30
 
- 
      Error Code:[0x8533] ERR COOLER «Main Cause» There is an error in the sample cooler's cooling unit. «Countermeasure» Turn power OFF and contact a Shimadzu sales or service representative. Show Detail - No:3585
- Date of Issue:2020/12/02 16:30
 
- 
      Error Code:[0x8530] CBM-20A ERROR «Main Cause» The injection volume exceeds the set value for the maximum volume of injection. «Countermeasure» Reduce the injection volume to a value less th... Show Detail - No:3584
- Date of Issue:2020/12/02 16:30
 
- 
      Error Code:[0x8529] SYSTEM ERROR xxx «Main Cause» There is an error in the Autosampler's internal circuits. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service represen... Show Detail - No:3583
- Date of Issue:2020/12/02 16:30
 
- 
      Error Code:[0x8513] ERR LPV HOME «Main Cause» Low-pressure valve's rotational position incorrect positions. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service repr... Show Detail - No:3582
- Date of Issue:2020/12/02 16:30
 
221-230 of Total 299