Error Code:[0x5264] ERROR PFILE CALC OVER
«Main Cause» The result of a calculation command in the pretreatment program exceeded the maximum value 65536. «Countermeasure» Revise the pretreatment prog... Show Detail
- No:3981
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x5265] ERROR PFILE ASSIGN RVAR
«Main Cause» The value substituted for a variable in a calculation command in the pretreatment program exceeded the maximum value 65536. «Countermeasure» Re... Show Detail
- No:3982
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x5266] ERROR PFILE ASSIGN LVAR
«Main Cause» The variable assigned by an assignment statement in the pretreatment program is not correct. «Countermeasure» Enter the correct variable. Show Detail
- No:3983
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x5267] ERROR PFILE IF OPERATOR
«Main Cause» The variable specified by an IF command in the pretreatment program is incorrect. «Countermeasure» Enter the correct variable. Show Detail
- No:3984
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x5268] ERROR PFILE NO ENDIF
«Main Cause» The ENDIF command corresponding to an IF command in the pretreatment program was not found. «Countermeasure» Add an ENDIF command. Show Detail
- No:3985
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x5269] ERROR PCG DISCONNECT
«Main Cause» Communication failed between the plate changer and autosampler during analysis. «Countermeasure» Check whether the plate changer power is switc... Show Detail
- No:3986
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x526A] ERROR PCG THETA HOME
«Main Cause» Something is obstructing the arm's line of movement or there is mechanical problem in the drive unit. «Countermeasure» Check for any obstructio... Show Detail
- No:3987
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x526B] ERROR PCG THETA SLIP
«Main Cause» Something is obstructing the arm's line of movement or there is mechanical problem in the drive unit. «Countermeasure» Check for any obstructio... Show Detail
- No:3988
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x526C] ERROR PCG Z HOME
«Main Cause» Something is obstructing the arm's line of movement or there is mechanical problem in the drive unit. «Countermeasure» Check for any obstructio... Show Detail
- No:3989
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x526D] ERROR PCG Z SLIP
«Main Cause» Something is obstructing the arm's line of movement or there is mechanical problem in the drive unit. «Countermeasure» Check for any obstructio... Show Detail
- No:3990