• Font Size
  • S
  • M
  • L

Frequently Referenced FAQ

FAQ within [ Autosampler ]

211-220 of Total 297

22 / 30page
  • Error Code:[0x8550] ERR LEAK DETECT

    «Main Cause» (1) Leak sensor was detected leakage. (2) Sensor is too sensitive. «Countermeasure» (1) Inspect the leakage from the flow line. Wipe out spille... Show Detail

    • No:3599
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x8548] STOP COOLER R

    «Main Cause» Since sample cooler R has operated without sample rack for more than 30 minutes, sample cooler R is stopped. «Countermeasure» Insert a sample r... Show Detail

    • No:3598
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x8547] STOP COOLER M

    «Main Cause» Since sample cooler M has operated without sample rack for more than 30 minutes, sample cooler M is stopped. «Countermeasure» Insert a sample r... Show Detail

    • No:3597
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x8546] STOP COOLER L

    «Main Cause» Since sample cooler L has operated without sample rack for more than 30 minutes, sample cooler L is stopped. «Countermeasure» Insert a sample r... Show Detail

    • No:3596
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x8544] NO RACK ID

    «Main Cause» Rack ID is not set correctly. «Countermeasure» Set a rack ID correctly. Show Detail

    • No:3595
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x8543] NOT ADJUSTED-R

    «Main Cause» A sample rack of which rack position has not been adjusted is inserted in rack plate R. «Countermeasure» Adjust the rack position. Show Detail

    • No:3594
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x8542] NOT ADJUSTED-M

    «Main Cause» A sample rack of which rack position has not been adjusted is inserted in rack plate M. «Countermeasure» Adjust the rack position. Show Detail

    • No:3593
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x8541] NOT ADJUSTED-L

    «Main Cause» A sample rack of which rack position has not been adjusted is inserted in rack plate L. «Countermeasure» Adjust the rack position. Show Detail

    • No:3592
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x8540] NOT ADJUST RACK

    «Main Cause» Analysis is started on the sample rack of which rack position has not been adjusted. «Countermeasure» Adjust the rack position. Show Detail

    • No:3591
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x8539] ERR TEMP SENS R

    «Main Cause» Sample cooler R temperature sensor error. «Countermeasure» Turn off the power and contact a Shimadzu sales or service representative. Show Detail

    • No:3590
    • Date of Issue:2020/12/02 16:30

211-220 of Total 297

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.