FAQ within [ Autosampler ]

191-200 of Total 297

20 / 30page
  • Error Code:[0x85EA] ERR RC HEATER

    «Main Cause» There is an error in the sample cooler's heating unit. «Countermeasure» Turn power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3619
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x85E8] RC FULL RACK

    «Main Cause» While a rack is being analyzed, another rack is set in its rack position. «Countermeasure» Remove the illegal rack or change it to the condensa... Show Detail

    • No:3618
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x85E0] RC NO RACK

    «Main Cause» There was no rack in the position specified for analysis and so the scheduled analysis was stopped. «Countermeasure» Set a rack in a position w... Show Detail

    • No:3617
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x85DD] ERR RC COOLER

    «Main Cause» Abnormality in cooler control unit. «Countermeasure» Turn power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3616
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x85DC] ERR RC TEMP SENS

    «Main Cause» There is an error in the sample cooler's temperature sensor. «Countermeasure» Turn power OFF and contact a Shimadzu sales or service representa... Show Detail

    • No:3615
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x85DB] ERR SET RACK

    «Main Cause» Error occurred while rack was being set. «Countermeasure» Turn power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3614
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x85DA] ERR CATCH RACK

    «Main Cause» An error occurred while a rack is being taken out. The plate has fallen while it is being loaded or dew condensation water gathers in large quan... Show Detail

    • No:3613
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x85D9] ERR RELEASE RACK

    «Main Cause» Rack was not released correctly. «Countermeasure» Turn power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3612
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x85D6] ERR HAND HOME Y

    «Main Cause» Motor's Y-axis movement (forward/backward) is incorrect. «Countermeasure» Turn power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3611
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x85D5] ERR HAND HOME Z

    «Main Cause» Motor's Z-axis movement (i.e., moving the rack vertically) is incorrect. «Countermeasure» Turn power OFF and contact a Shimadzu sales or servic... Show Detail

    • No:3610
    • Date of Issue:2020/12/02 16:30