Frequently Referenced FAQ

FAQ within [ Autosampler ]

181-190 of Total 297

19 / 30page
  • Error Code:[0x001D]

    «Main Cause» A vial number specified in the sequence is out of range. «Countermeasure» Specify vial numbers that correspond to the given rack. Show Detail

    • No:3824
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x0019]

    «Main Cause» The autosampler was unable to recognize the rack or the rack is not located in the autosampler. «Countermeasure» Check the rack. Show Detail

    • No:3823
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x000F]

    «Main Cause» A vial number specified in the sequence is out of range. «Countermeasure» Specify vial numbers that correspond to the given rack. Show Detail

    • No:3822
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x8568]

    «Main Cause» The remaining volume of rinse solution is less than the check level. «Countermeasure» Replace the rinse solution or reset the rinse solution vo... Show Detail

    • No:3814
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x85F4] ERR DISCONNECT

    «Main Cause» The communication between the instrument and the autosampler is cut off due to a communication error. «Countermeasure» Re-insert both connector... Show Detail

    • No:3625
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x85F1] REMOVE RACK

    «Main Cause» The same rack has been set twice. After analysis is performed for this rack, [RC FULL RACK] error is displayed and the scheduled operation is st... Show Detail

    • No:3624
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x85EF] SET RACK

    «Main Cause» There is a stack for which a rack has not been set. When analysis is performed for this rack, [RC NO. RACK] error is displayed and the scheduled... Show Detail

    • No:3623
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x85EE] COOLER ON AGAIN

    «Main Cause» The stack was closed and so the sample cooler turned ON. «Countermeasure» - Show Detail

    • No:3622
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x85ED] COOLER OFF

    «Main Cause» A stack was left open for more than 30 minutes and so the sample cooler has been turned OFF. «Countermeasure» Close the stack to turn the samp... Show Detail

    • No:3621
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x85EC] CLOSE STACK

    «Main Cause» A stack was left open for more than 30 minutes and so the sample cooler has been turned OFF. «Countermeasure» Close the stack. Show Detail

    • No:3620
    • Date of Issue:2020/12/02 16:30