Error Code:[0x523C] ERROR PFILE ASPIR SPEED
«Main Cause» The suction speed specified for suction in the pretreatment program exceeds the setting range. «Countermeasure» Change the setting to a value w... Show Detail
- No:3941
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x523D] ERROR PFILE DISP VOLUME
«Main Cause» The liquid volume specified for discharge in the pretreatment program exceeds the setting range. «Countermeasure» Change the setting to a value... Show Detail
- No:3942
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x523E] ERROR PFILE DISP SPEED
«Main Cause» The discharge speed specified for discharge in the pretreatment program exceeds the setting range. «Countermeasure» Change the setting to a val... Show Detail
- No:3943
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x523F] ERROR PFILE AIR.A VOL
«Main Cause» The liquid volume specified for air intake in the pretreatment program exceeds the setting range. «Countermeasure» Change the setting to a valu... Show Detail
- No:3944
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x5240] ERROR PFILE AIR.A SPEED
«Main Cause» The suction speed specified for air intake in the pretreatment program exceeds the setting range. «Countermeasure» Change the setting to a valu... Show Detail
- No:3945
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x5241] ERROR PFILE MIX POS
«Main Cause» The needle is not positioned inside or above a sample vial when the MIX command is executed in the pretreatment program. «Countermeasure» Add a... Show Detail
- No:3946
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x5242] ERROR PFILE MIX COUNT
«Main Cause» The number of repetitions specified by the MIX command in the pretreatment program exceeds the setting range. «Countermeasure» Change the setti... Show Detail
- No:3947
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x5243] ERROR PFILE MIX AIR VOL
«Main Cause» The upper/lower air volume specified by the MIX command in the pretreatment program exceeds the setting range. «Countermeasure» Change the sett... Show Detail
- No:3948
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x5244] ERROR PFILE MIX VOLUME
«Main Cause» A suction volume specified by the MIX command in the pretreatment program exceeds the setting range. «Countermeasure» Change the setting to a v... Show Detail
- No:3949
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x5245] ERROR PFILE MIX A.SPEED
«Main Cause» The suction speed specified by the MIX command in the pretreatment program exceeds the setting range. «Countermeasure» Change the setting to a ... Show Detail
- No:3950
- Date of Issue:2020/12/02 16:44