Error Code:[0x5206] ERROR X SLIP
«Main Cause» Something is obstructing the X-direction (left-right) line of movement or there is a mechanical problem in the drive unit. «Countermeasure» Che... Show Detail
- No:3887
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x5205] ERROR X HOME POSITION
«Main Cause» Something is obstructing the needle's line of movement or there is a mechanical problem in the drive unit. «Countermeasure» Check for any obstr... Show Detail
- No:3886
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x5204] FATAL NOT CONNECT SENS
«Main Cause» Communication with the plate reading board failed. «Countermeasure» Contact a Shimadzu sales or service representative. Show Detail
- No:3885
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x5203] FATAL NOT CONNECT PLB
«Main Cause» Communication with the measuring pump and low pressure valve control board failed. «Countermeasure» Contact a Shimadzu sales or service represe... Show Detail
- No:3884
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x5202] FATAL NOT CONNECT XZ
«Main Cause» Communication with the X-axis and Z-axis motor control board failed. «Countermeasure» Contact a Shimadzu sales or service representative. Show Detail
- No:3883
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x5201] FATAL NOT PROTECTED
«Main Cause» The configuration settings failed to be retrieved. «Countermeasure» If an error occurs each time started, contact a Shimadzu sales or service r... Show Detail
- No:3882
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x5200] FATAL SYSTEM ERROR
«Main Cause» A system error occurred. «Countermeasure» Switch the power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:3881
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x0061] ERROR LEAK DETECTED
«Main Cause» Leakage occurred from a tubing joint or valve gap. «Countermeasure» Check the following locations for leakage. ・If there is leakage from the in... Show Detail
- No:3880
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x0018] ERROR PFILE NOT EXIST
«Main Cause» The pretreatment program size exceeded the limit value (max. 255 bytes per file or 4,096 bytes for the entire program). «Countermeasure» Optimi... Show Detail
- No:3879
- Date of Issue:2020/12/02 16:44
-
«Main Cause» The injection volume exceeded the [injection volume upper limit] setting value. «Countermeasure» Inject a volume that does not exceed the [inje... Show Detail
- No:3831
- Date of Issue:2020/12/02 16:43