• Font Size
  • S
  • M
  • L

FAQ within [ Autosampler ]

161-170 of Total 297

17 / 30page
  • Error Code:[0x5210] ERROR LPV DATA

    «Main Cause» The low pressure valve is disconnected or there is an electrical contact problem. «Countermeasure» Check whether the low pressure valve is disc... Show Detail

    • No:3897
    • Date of Issue:2020/12/02 16:44
  • Error Code:[0x520F] ERROR LPV SLIP

    «Main Cause» The torque required to rotate the valve increased or there is a mechanical problem in the drive unit. «Countermeasure» Click the [CE] button an... Show Detail

    • No:3896
    • Date of Issue:2020/12/02 16:44
  • Error Code:[0x520E] ERROR LPV HOME POSITION

    «Main Cause» The torque required to rotate the valve increased or there is a mechanical problem in the drive unit. «Countermeasure» Click the [CE] button an... Show Detail

    • No:3895
    • Date of Issue:2020/12/02 16:44
  • Error Code:[0x520D] ERROR HPV TYPE

    «Main Cause» The type of high pressure valve is not correct. «Countermeasure» Install a 6-port 2-position valve. Show Detail

    • No:3894
    • Date of Issue:2020/12/02 16:44
  • Error Code:[0x520C] ERROR HPV DATA

    «Main Cause» The high pressure valve is disconnected or there is an electrical contact problem. «Countermeasure» Check whether the high pressure valve is di... Show Detail

    • No:3893
    • Date of Issue:2020/12/02 16:44
  • Error Code:[0x520B] ERROR HPV SLIP

    «Main Cause» The torque required to rotate the valve increased or there is a mechanical problem in the drive unit. «Countermeasure» Click the [CE] button an... Show Detail

    • No:3892
    • Date of Issue:2020/12/02 16:44
  • Error Code:[0x520A] ERROR HPV HOME POSITION

    «Main Cause» The torque required to rotate the valve increased or there is a mechanical problem in the drive unit. «Countermeasure» Click the [CE] button an... Show Detail

    • No:3891
    • Date of Issue:2020/12/02 16:44
  • Error Code:[0x5209] FATAL Z HOME POSITION

    «Main Cause» Something is obstructing the needle's line of movement or there is a mechanical problem in the drive unit. «Countermeasure» Check for any obstr... Show Detail

    • No:3890
    • Date of Issue:2020/12/02 16:44
  • Error Code:[0x5208] ERROR Y SLIP

    «Main Cause» Something is obstructing the Y-direction (front-rear) line of movement or there is a mechanical problem in the drive unit. «Countermeasure» Che... Show Detail

    • No:3889
    • Date of Issue:2020/12/02 16:44
  • Error Code:[0x5207] ERROR Y HOME POSITION

    «Main Cause» Something is obstructing the needle's line of movement or there is a mechanical problem in the drive unit. «Countermeasure» Check for any obstr... Show Detail

    • No:3888
    • Date of Issue:2020/12/02 16:44

161-170 of Total 297

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.