  • Error Code:[0x521A] FATAL INNER FAN

    «Main Cause» A fan rotation problem was detected for the fan inside the cabinet. «Countermeasure» Check for foreign matter caught in the fan in the cabinet.... Show Detail

    • No:3907
    • Date of Issue:2020/12/02 16:44
  • Error Code:[0x5219] FATAL PELTIER2 VOLTAGE

    «Main Cause» The Peltier element used for temperature adjustment is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a... Show Detail

    • No:3906
    • Date of Issue:2020/12/02 16:44
  • Error Code:[0x5218] FATAL PELTIER1 VOLTAGE

    «Main Cause» The Peltier element used for temperature adjustment is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a... Show Detail

    • No:3905
    • Date of Issue:2020/12/02 16:44
  • Error Code:[0x5217] ERROR TEMP SENSOR2

    «Main Cause» Temperature sensor 2 inside the cabinet detected a temperature outside the specified range. The temperature sensor malfunctioned or the temperat... Show Detail

    • No:3904
    • Date of Issue:2020/12/02 16:44
  • Error Code:[0x5216] ERROR TEMP SENSOR1

    «Main Cause» Temperature sensor 1 inside the cabinet detected a temperature outside the specified range. The temperature sensor malfunctioned or the temperat... Show Detail

    • No:3903
    • Date of Issue:2020/12/02 16:44
  • Error Code:[0x5215] FATAL NEEDLE PROTECTED

    «Main Cause» The needle cannot be lowered because an obstruction other than the sample vial or injection port was detected. «Countermeasure» Switch the inst... Show Detail

    • No:3902
    • Date of Issue:2020/12/02 16:44
  • Error Code:[0x5214] ERROR PUMP SLIP

    «Main Cause» Plunger wear has increased or there is a problem in the drive unit. «Countermeasure» Click the [CE] button and try again. If the error is not r... Show Detail

    • No:3901
    • Date of Issue:2020/12/02 16:44
  • Error Code:[0x5213] ERROR ADJUST PUMP

    «Main Cause» Plunger wear has increased or there is a problem in the drive unit. «Countermeasure» Click the [CE] button and try again. If the error is not r... Show Detail

    • No:3900
    • Date of Issue:2020/12/02 16:44
  • Error Code:[0x5212] ERROR PUMP HOME POS

    «Main Cause» Plunger wear has increased or there is a problem in the drive unit. «Countermeasure» Click the [CE] button and try again. If the error is not r... Show Detail

    • No:3899
    • Date of Issue:2020/12/02 16:44
  • Error Code:[0x5211] ERROR LPV TYPE

    «Main Cause» The type of low pressure valve is incorrect. «Countermeasure» Install a 7-port valve. Show Detail

    • No:3898
    • Date of Issue:2020/12/02 16:44