• Font Size
  • S
  • M
  • L

FAQ within [ Autosampler ]

141-150 of Total 297

15 / 30page
  • Error Code:[0x5224] ERROR NO VIAL

    «Main Cause» No sample vial is placed at the vial number position specified in the sequence. «Countermeasure» Place a sample vial at the vial number positio... Show Detail

    • No:3917
    • Date of Issue:2020/12/02 16:44
  • Error Code:[0x5223] ERROR LINK TIMEOUT

    «Main Cause» The connection to the system controller was disconnected during an analysis. «Countermeasure» Check whether the optical link cable is disconnec... Show Detail

    • No:3916
    • Date of Issue:2020/12/02 16:44
  • Error Code:[0x5222] FATAL Y ENCODER

    «Main Cause» The encoder sensor is not counting normally in the needle Y-axis (front-rear) direction. «Countermeasure» Click the [CE] button and try again. ... Show Detail

    • No:3915
    • Date of Issue:2020/12/02 16:44
  • Error Code:[0x5221] FATAL X ENCODER

    «Main Cause» The encoder sensor is not counting normally in the needle X-axis (left-right) direction. «Countermeasure» Click the [CE] button and try again. ... Show Detail

    • No:3914
    • Date of Issue:2020/12/02 16:44
  • Error Code:[0x5220] FATAL LEAK SENSOR

    «Main Cause» The leak sensor cannot be detected properly. The leak sensor might not be connected. «Countermeasure» Switch the instrument power OFF and conta... Show Detail

    • No:3913
    • Date of Issue:2020/12/02 16:44
  • Error Code:[0x521F] FATAL LEAK SENSOR

    «Main Cause» The leak sensor cannot be detected properly. A broken circuit may have occurred in the leak sensor. «Countermeasure» Switch the instrument powe... Show Detail

    • No:3912
    • Date of Issue:2020/12/02 16:44
  • Error Code:[0x521E] ERROR LEAK DETECTED

    «Main Cause» Leakage occurred from a tubing joint or valve gap. «Countermeasure» Check the following locations for leakage. ・If there is leakage from the in... Show Detail

    • No:3911
    • Date of Issue:2020/12/02 16:44
  • Error Code:[0x521D] WARNING PELTIER STOP

    «Main Cause» No sample rack was detected for a given period. «Countermeasure» Place the sample rack and start temperature control. Show Detail

    • No:3910
    • Date of Issue:2020/12/02 16:44
  • Error Code:[0x521C] FATAL POWER FAN

    «Main Cause» A fan rotation problem was detected in the power supply fan. «Countermeasure» Check for foreign matter caught in the power supply fan. If that ... Show Detail

    • No:3909
    • Date of Issue:2020/12/02 16:44
  • Error Code:[0x521B] FATAL OUTER FAN

    «Main Cause» A rotation problem was detected for the fan on the back of the instrument. «Countermeasure» Check for foreign matter caught in the fan on the b... Show Detail

    • No:3908
    • Date of Issue:2020/12/02 16:44

141-150 of Total 297

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.