Error Code:[0x5238] ERROR PFILE ILLEGAL VAL
«Main Cause» A variable not present in the pretreatment program is specified. «Countermeasure» Check the variable. Show Detail
- No:3937
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x5237] ERROR PFILE ILLEGAL CMD
«Main Cause» A command not present in the pretreatment program is specified. «Countermeasure» Check the command. Show Detail
- No:3936
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x5236] ERROR MISMATCH PLATE
«Main Cause» The sample plate type does not match the type specified in the method file or the barcode label attached to the sample plate does not match the ... Show Detail
- No:3935
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x5235] ERROR DETECT CLOGGING
«Main Cause» A blockage was detected in the high pressure flow channels inside the autosampler. «Countermeasure» Perform auto-purge. Show Detail
- No:3934
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x5234] WARNING DETECT CLOGGING
«Main Cause» A blockage was detected in the high pressure flow channels inside the autosampler. «Countermeasure» Perform auto-purge. Show Detail
- No:3933
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x5233] ERROR PFILE
«Main Cause» The pretreatment program size exceeded the limit value (max. 255 bytes per file or 4,096 bytes for the entire program). «Countermeasure» Optimi... Show Detail
- No:3932
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x5232] ERROR INJ VOLUME
«Main Cause» The injection volume exceeds the maximum injection volume setting value. «Countermeasure» Change the injection volume to a value less than the ... Show Detail
- No:3931
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x5231] ERROR VIAL NUMVER
«Main Cause» The specified vial number exceeds the number of vials that can be placed on the sample plate. «Countermeasure» Change the vial number setting. Show Detail
- No:3930
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x5230] WARNING SYRINGE
«Main Cause» Syringe use was specified in autosampler settings, but no syringe is connected. «Countermeasure» Connect a syringe to the autosampler and switc... Show Detail
- No:3929
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x522F] WARNING PLATE ID
«Main Cause» The type of sample plate placed on the sample rack was not read correctly. «Countermeasure» Reposition the sample plate. Show Detail
- No:3928
- Date of Issue:2020/12/02 16:44