«Main Cause» There is a stack for which a rack has not been set. When analysis is performed for this rack, [RC NO. RACK] error is displayed and the scheduled... Show Detail
- No:3623
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x85F1] REMOVE RACK
«Main Cause» The same rack has been set twice. After analysis is performed for this rack, [RC FULL RACK] error is displayed and the scheduled operation is st... Show Detail
- No:3624
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x85F4] ERR DISCONNECT
«Main Cause» The communication between the instrument and the autosampler is cut off due to a communication error. «Countermeasure» Re-insert both connector... Show Detail
- No:3625
- Date of Issue:2020/12/02 16:30
-
«Main Cause» The remaining volume of rinse solution is less than the check level. «Countermeasure» Replace the rinse solution or reset the rinse solution vo... Show Detail
- No:3814
- Date of Issue:2020/12/02 16:43
-
«Main Cause» A vial number specified in the sequence is out of range. «Countermeasure» Specify vial numbers that correspond to the given rack. Show Detail
- No:3822
- Date of Issue:2020/12/02 16:43
-
«Main Cause» The autosampler was unable to recognize the rack or the rack is not located in the autosampler. «Countermeasure» Check the rack. Show Detail
- No:3823
- Date of Issue:2020/12/02 16:43
-
«Main Cause» A vial number specified in the sequence is out of range. «Countermeasure» Specify vial numbers that correspond to the given rack. Show Detail
- No:3824
- Date of Issue:2020/12/02 16:43
-
«Main Cause» The injection volume exceeded the [injection volume upper limit] setting value. «Countermeasure» Inject a volume that does not exceed the [inje... Show Detail
- No:3831
- Date of Issue:2020/12/02 16:43
-
Error Code:[0x0018] ERROR PFILE NOT EXIST
«Main Cause» The pretreatment program size exceeded the limit value (max. 255 bytes per file or 4,096 bytes for the entire program). «Countermeasure» Optimi... Show Detail
- No:3879
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x0061] ERROR LEAK DETECTED
«Main Cause» Leakage occurred from a tubing joint or valve gap. «Countermeasure» Check the following locations for leakage. ・If there is leakage from the in... Show Detail
- No:3880
- Date of Issue:2020/12/02 16:44