• Font Size
  • S
  • M
  • L

Frequently Referenced FAQ

FAQ within [ Autosampler ]

111-120 of Total 299

≪ 12 / 30page ≫
  • Error Code:[0x85EF] SET RACK

    «Main Cause» There is a stack for which a rack has not been set. When analysis is performed for this rack, [RC NO. RACK] error is displayed and the scheduled... Show Detail

    • No:3623
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x85F1] REMOVE RACK

    «Main Cause» The same rack has been set twice. After analysis is performed for this rack, [RC FULL RACK] error is displayed and the scheduled operation is st... Show Detail

    • No:3624
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x85F4] ERR DISCONNECT

    «Main Cause» The communication between the instrument and the autosampler is cut off due to a communication error. «Countermeasure» Re-insert both connector... Show Detail

    • No:3625
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x8568]

    «Main Cause» The remaining volume of rinse solution is less than the check level. «Countermeasure» Replace the rinse solution or reset the rinse solution vo... Show Detail

    • No:3814
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x000F]

    «Main Cause» A vial number specified in the sequence is out of range. «Countermeasure» Specify vial numbers that correspond to the given rack. Show Detail

    • No:3822
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x0019]

    «Main Cause» The autosampler was unable to recognize the rack or the rack is not located in the autosampler. «Countermeasure» Check the rack. Show Detail

    • No:3823
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x001D]

    «Main Cause» A vial number specified in the sequence is out of range. «Countermeasure» Specify vial numbers that correspond to the given rack. Show Detail

    • No:3824
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x856A]

    «Main Cause» The injection volume exceeded the [injection volume upper limit] setting value. «Countermeasure» Inject a volume that does not exceed the [inje... Show Detail

    • No:3831
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x0018] ERROR PFILE NOT EXIST

    «Main Cause» The pretreatment program size exceeded the limit value (max. 255 bytes per file or 4,096 bytes for the entire program). «Countermeasure» Optimi... Show Detail

    • No:3879
    • Date of Issue:2020/12/02 16:44
  • Error Code:[0x0061] ERROR LEAK DETECTED

    «Main Cause» Leakage occurred from a tubing joint or valve gap. «Countermeasure» Check the following locations for leakage. ・If there is leakage from the in... Show Detail

    • No:3880
    • Date of Issue:2020/12/02 16:44

111-120 of Total 299

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Technology Inc.