Keyword Search
Frequently Referenced FAQ
FAQ within [ Autosampler ]
- 
      Error Code:[0x524E] ERROR PFILE GOTO FILE # «Main Cause» The file number specified by the GOTO command in the pretreatment program exceeds the setting range. «Countermeasure» Change the setting to a v... Show Detail - No:3959
- Date of Issue:2020/12/02 16:44
 
- 
      Error Code:[0x524D] ERROR PFILE WAIT.M VAR «Main Cause» The wait time specified by the pretreatment program exceeds the setting range. «Countermeasure» Change the setting to a value within the specif... Show Detail - No:3958
- Date of Issue:2020/12/02 16:44
 
- 
      Error Code:[0x524C] ERROR PFILE NSTRK RANGE «Main Cause» The needle stroke distance specified by the pretreatment program exceeds the setting range. «Countermeasure» Change the setting to a value with... Show Detail - No:3957
- Date of Issue:2020/12/02 16:44
 
- 
      Error Code:[0x524B] ERROR PFILE NSTRK POS «Main Cause» The needle is not positioned within or above a sample vial. «Countermeasure» Add a VIAL.N or other command to move the needle to the sample vial. Show Detail - No:3956
- Date of Issue:2020/12/02 16:44
 
- 
      Error Code:[0x524A] ERROR PFILE APPORT SPD «Main Cause» A suction speed specified by the APPORT command in the pretreatment program exceeds the setting range. «Countermeasure» Change the setting to a... Show Detail - No:3955
- Date of Issue:2020/12/02 16:44
 
- 
      Error Code:[0x5249] ERROR PFILE APPORT VOL «Main Cause» A sample volume specified by the APPORT command in the pretreatment program exceeds the setting range. «Countermeasure» Change the setting to a... Show Detail - No:3954
- Date of Issue:2020/12/02 16:44
 
- 
      Error Code:[0x5248] ERROR PFILE APPORT TO «Main Cause» The vial number specified by the APPORT command in the pretreatment program exceeds the setting range. «Countermeasure» Change the setting to a... Show Detail - No:3953
- Date of Issue:2020/12/02 16:44
 
- 
      Error Code:[0x5247] ERROR PFILE APPORT FROM «Main Cause» The vial number specified by the APPORT command in the pretreatment program exceeds the setting range. «Countermeasure» Change the setting to a... Show Detail - No:3952
- Date of Issue:2020/12/02 16:44
 
- 
      Error Code:[0x5246] ERROR PFILE MIX D.SPEED «Main Cause» The discharge speed specified by the MIX command in the pretreatment program exceeds the setting range. «Countermeasure» Change the setting to ... Show Detail - No:3951
- Date of Issue:2020/12/02 16:44
 
- 
      Error Code:[0x5245] ERROR PFILE MIX A.SPEED «Main Cause» The suction speed specified by the MIX command in the pretreatment program exceeds the setting range. «Countermeasure» Change the setting to a ... Show Detail - No:3950
- Date of Issue:2020/12/02 16:44
 
101-110 of Total 299