FAQ within [ Autosampler ]
-
Error Code:[0x85DA] ERR CATCH RACK
«Main Cause» An error occurred while a rack is being taken out. The plate has fallen while it is being loaded or dew condensation water gathers in large quan... Show Detail
- No:3613
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x85DB] ERR SET RACK
«Main Cause» Error occurred while rack was being set. «Countermeasure» Turn power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:3614
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x85DC] ERR RC TEMP SENS
«Main Cause» There is an error in the sample cooler's temperature sensor. «Countermeasure» Turn power OFF and contact a Shimadzu sales or service representa... Show Detail
- No:3615
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x85DD] ERR RC COOLER
«Main Cause» Abnormality in cooler control unit. «Countermeasure» Turn power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:3616
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x85E0] RC NO RACK
«Main Cause» There was no rack in the position specified for analysis and so the scheduled analysis was stopped. «Countermeasure» Set a rack in a position w... Show Detail
- No:3617
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x85E8] RC FULL RACK
«Main Cause» While a rack is being analyzed, another rack is set in its rack position. «Countermeasure» Remove the illegal rack or change it to the condensa... Show Detail
- No:3618
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x85EA] ERR RC HEATER
«Main Cause» There is an error in the sample cooler's heating unit. «Countermeasure» Turn power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:3619
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x85EC] CLOSE STACK
«Main Cause» A stack was left open for more than 30 minutes and so the sample cooler has been turned OFF. «Countermeasure» Close the stack. Show Detail
- No:3620
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x85ED] COOLER OFF
«Main Cause» A stack was left open for more than 30 minutes and so the sample cooler has been turned OFF. «Countermeasure» Close the stack to turn the samp... Show Detail
- No:3621
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x85EE] COOLER ON AGAIN
«Main Cause» The stack was closed and so the sample cooler turned ON. «Countermeasure» - Show Detail
- No:3622
- Date of Issue:2020/12/02 16:30