(ICPE) Stopping the ICPE in Daily Use
1. Turn off the plasma. 2. Clean the inside of the stand and the glassware. 3. Waste liquid is treated. 4. After stopping the vacuum pump,... Show Detail
- No:6732
- Date of Issue:2021/07/30 09:13
- Last Updated:2021/10/04 09:46
- Category: Multi type(ICPE)
- Brached FAQ
-
(ICPE) Leak Valve Operation (After vacuum pump is stopped)
NOTE: When stopping the vacuum pump, release the molecular sieve case toatmospheereic pressure. 1.After stopping the vacuum pump, open the leak valve.... Show Detail
- No:6731
- Date of Issue:2021/07/30 09:07
- Category: Multi type(ICPE)
-
(ICPE) Starting the ICPE in Daily Use
1. Supply cooling water to the ICPE. If mains water is used, open the main valve of the mains water supply. If a cooling water circulation system is used... Show Detail
- No:6730
- Date of Issue:2021/07/30 09:03
- Last Updated:2023/12/27 15:49
- Category: Multi type(ICPE)
-
(ICPE) Cleaning the Axial-View Observation Orifice Assembly
There are ICPE-9800 Series Standard: 211-87758-42 Orifice L2 ASSY (formini torch ICPE-9800 Series Option: 211-88571-42 Orifice S2 ASSY (for optionalpla... Show Detail
- No:6729
- Date of Issue:2021/07/29 16:33
- Last Updated:2021/10/04 10:16
- Category: Multi type(ICPE)
-
(ICPE) Cleaning and Replacement of Reflect Mirror in Cool...
1. Loosen the mirror holder's set screws and remove the mirror holder from thejacket. 2. When cleaning the mirror, leave it on the mirror holder and c... Show Detail
- No:6728
- Date of Issue:2021/07/29 16:16
- Last Updated:2021/10/04 09:47
- Category: Multi type(ICPE)
- Brached FAQ
-
(ICPE) Attaching and Removing the Cooling Jacket
1. Remove the two cooling water tubes. 2. Remove the argon gas purge tube. 3. Remove the two attachment screws. NOTE: When the cool... Show Detail
- No:6727
- Date of Issue:2021/07/29 16:00
- Last Updated:2021/10/04 09:48
- Category: Multi type(ICPE)
- Brached FAQ
-
(ICPE) Replacing the Peristaltic Pump Tube
Since the sample pressure feed tube is constrained by the rotor and guide, it is subjectto plastic deformation and eventually samples will no longer feed thr... Show Detail
- No:6726
- Date of Issue:2021/07/29 14:31
- Category: Multi type(ICPE)
- Brached FAQ
-
(ICPS) Check and Replacement of Glassware
1) Holding the chamber, remove the ball joint clip. 2) Loosen the clamp of torch stand, pull out the torch downward and remove the tube connec... Show Detail
- No:6767
- Date of Issue:2021/07/29 14:03
- Category: Sequential type(ICPS)
- Brached FAQ
-
(ICPE) Replacing and Inspecting Glassware, Sample Introdu...
WARNING: There is a risk of burn injury. Do NOT touch the plasma standinterior for 3 minutes after the plasma is extinguished. CAUTION: Glass parts break ... Show Detail
- No:6724
- Date of Issue:2021/07/29 14:03
- Category: Multi type(ICPE)
- Brached FAQ
-
(ICPS) Maintenance of Vacuum Pump
The molecular sieve can be replaced by a new one with the spectrometerkept at a vacuum status. Parts number: 239-07003-08 Molecular siev Parts numb... Show Detail
- No:6775
- Date of Issue:2021/07/28 10:55
- Category: Sequential type(ICPS)
- Brached FAQ