FAQ - Frequently Asked Questions
- Products Search > Elemental Analysis > Inductively Coupled Plasma Emission Spectroscopy(ICP)
Keyword Search
Inductively Coupled Plasma Emission Spectroscopy(ICP)
Frequently Referenced FAQ
FAQ within [ Inductively Coupled Plasma Emission Spectroscopy(ICP) ]
-
(ICPE) Filter Cleaning and Replacement (Autosampler)
If the filter is clogged, remove the cover on the left side of the ASC-9800 and clean the filter. If the filter has deteriorated, replace it with a new filt... Show Detail
- No:6742
- Date of Issue:2021/08/03 10:56
- Category: Multi type(ICPE)
- Brached FAQ
-
(ICPE) Replacing the Peristaltic Pump Tubing (Autosampler)
If the flow rate of the peristaltic pump used to supply rinse solution decreases, replace the tubing in the peristaltic pump according to the following proce... Show Detail
- No:6741
- Date of Issue:2021/08/03 10:51
- Category: Multi type(ICPE)
- Brached FAQ
-
(ICPE) Mounting the Nozzle (Autosampler)
1. Insert the nozzle (Parts number: 206-50290-91) into the nozzle hole located on the arm. 2. Lower the nozzle tip by 103 mm from the bottom of the arm an... Show Detail
- No:6740
- Date of Issue:2021/08/03 10:21
- Category: Multi type(ICPE)
- Brached FAQ
-
(ICPE) The spectral intensity for elements is high even w...
• The blank liquid is dirty. • The torch is dirty. • The chamber is dirty. Click here for "Replacing and Inspecting Glassware, Sample Introduction Tub... Show Detail
- No:6739
- Date of Issue:2021/07/30 10:11
- Category: Multi type(ICPE)
-
(ICPE) Condensation occurs on the RF coil or the cooling ...
• Wipe off the water droplets with gauze or some otherappropriate material and ignite the plasma. • Checks that the installation location meets the installa... Show Detail
- No:6738
- Date of Issue:2021/07/30 10:11
- Last Updated:2021/10/04 10:07
- Category: Multi type(ICPE)
-
(ICPE) The spectral intensity is unstable
• The ICPEinstallation location does not meetrequirements related to installation location or facilities. • One of the causes listed for problem "The spectr... Show Detail
- No:6737
- Date of Issue:2021/07/30 10:11
- Category: Multi type(ICPE)
-
(ICPE) The spectral intensity dropped significantly
• Deposits adhered to the tip of the nebulizer. • Deterioration of nebulizer • Deterioration of sample introduction tube • Bubbles entered the sample intr... Show Detail
- No:6736
- Date of Issue:2021/07/30 10:09
- Category: Multi type(ICPE)
-
(ICPE) Cleaning/Replacing the Purge Tube(ICPE-9820)
When a high concentration sample is analyzed, some of the sample may adhere to thepurge tube. If the tube becomes heavily contaminated, remove the purge tube... Show Detail
- No:6735
- Date of Issue:2021/07/30 09:57
- Category: Multi type(ICPE)
-
(ICPE) Cleaning the Entrance Window
WARNING:There is a risk of burn injury.Do NOT start work inside the plasma stand for 3 minutes afterthe plasma is extinguished. NOTE: If it takes much... Show Detail
- No:6734
- Date of Issue:2021/07/30 09:52
- Category: Multi type(ICPE)
- Brached FAQ
-
1. Recovery after Emergency Stop After stopping the ICPEby turning OFF the power switch, perform the followingrecovery procedure. 1) Turn OFF the comp... Show Detail
- No:6733
- Date of Issue:2021/07/30 09:19
- Category: Multi type(ICPE)
21-30 of Total 45