FAQ - Frequently Asked Questions
- Products Search > Elemental Analysis > Inductively Coupled Plasma Emission Spectroscopy(ICP)
Keyword Search
Inductively Coupled Plasma Emission Spectroscopy(ICP)
Frequently Referenced FAQ
FAQ within [ Inductively Coupled Plasma Emission Spectroscopy(ICP) ]
-
1.Turning off of plasma 1) Inspire pure water for three to five minutes. This allows the sample suction tube, nebulizer, chamber and torch to be cle... Show Detail
- No:6780
- Date of Issue:2021/08/18 16:14
- Category: Sequential type(ICPS)
- Brached FAQ
-
(ICPS) Starting Up of Equipment
1.Supply of power to equipment 1) Turn ON the main breaker B1 (lift the lever). The pilot lamp S2is lit. 2) Turn ON (Press) the pump start switc... Show Detail
- No:6778
- Date of Issue:2021/08/18 16:10
- Category: Sequential type(ICPS)
- Brached FAQ
-
(ICPS) Replacement of RF Power Fuse
When the RF power generator fuse (F1 & F2) is blown, replace it following the procedures below. WARNING:You may get an electric shock.. Before starting oper... Show Detail
- No:6777
- Date of Issue:2021/08/18 15:57
- Category: Sequential type(ICPS)
- Brached FAQ
-
(ICPS) Cleaning of Condensing Lens
If the condensing lens is used for a long time, its surface is contaminated, deteriorating transmittance. 1) Remove the stand top cover. 2) Pull... Show Detail
- No:6774
- Date of Issue:2021/08/18 15:30
- Category: Sequential type(ICPS)
- Brached FAQ
-
Check the tank periodically to make sure the following. ・ The waste liquid is not overflow. ・ The drain tube does not come in contact with the liquid surfa... Show Detail
- No:6772
- Date of Issue:2021/08/18 15:08
- Last Updated:2021/09/13 09:15
- Category: Sequential type(ICPS)
-
Approximately 10L of pure water is used as the coil cooling water. Replace the water withnew one when fur forms in it. WARNIGN:You may get an electric... Show Detail
- No:6771
- Date of Issue:2021/08/18 15:08
- Category: Sequential type(ICPS)
-
(ICPS) Cleaning of Orifice for Axial Observation(ICPS-7510)
NOTE: Cleaning the orifice prevents contamination of a sample and abnormaldischarge. WARNING:Never touch the inside of the stand for three minutes after e... Show Detail
- No:6769
- Date of Issue:2021/08/18 15:05
- Last Updated:2022/01/17 15:47
- Category: Sequential type(ICPS)
-
(ICPS) Maintenance in the plasma stand(ICPS-7510)
1.Cleaning part (1) Sheet copper connected with RF coil and the connecting part (2) Connecting part of this sheet copper and matching-Box.And the circumf... Show Detail
- No:6770
- Date of Issue:2021/08/18 15:04
- Category: Sequential type(ICPS)
-
(ICPS) Cleaning of Folding Mirror for Axial Observation(I...
WARNING:You may get a burn. Before starting the procedures below, turn OFF theplasma and wait until the water cooled jacket is cooled sufficiently (at least3... Show Detail
- No:6768
- Date of Issue:2021/08/18 14:50
- Category: Sequential type(ICPS)
-
(ICPS) Operation Check of Safety Device
WARNING: You may a burn. Don't touch the exhaust pipe on a plasma stand, or don't look in. 1) Argon gas pressure switch If you stop supplying argo... Show Detail
- No:6766
- Date of Issue:2021/08/18 12:48
- Category: Sequential type(ICPS)
- Brached FAQ
11-20 of Total 45