PURn calc. prss out of range[Error Code:5042 5043 5044 50...
Set the purge flow within the AFC control range. Show Detail
- No:5998
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 10:14
- Category: Error/Warning
-
DET#n FTD overcurrent detected[Error Code:4105 4106 4107 ...
Turn off the main power of the instrument, and confirm the following items. The set value of the current exceeds the maximum operating current. The mak... Show Detail
- No:5997
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 10:12
- Category: Error/Warning
-
DET#n TCD overcurrent detected[Error Code:4101 4102 4103 ...
Turn off the main power of the instrument, and confirm the following items. The set value of the current exceeds the maximum operating current. The mak... Show Detail
- No:5996
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 10:07
- Category: Error/Warning
-
Flow# command stopped due to error[Error Code:4021 4022 4...
When any of these errors occurs, another error should have occurred regarding the flow controller. Check the actions necessary for the error. Show Detail
- No:5995
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 10:01
- Category: Error/Warning
-
AOC link error[Error Code:4013]
Turn off the AOC power, reconnect the cable, and turn on the AOC again. If the error appears even after checking the above, turn off the main power of the in... Show Detail
- No:5994
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 09:53
- Category: Error/Warning
-
AOC2 transmission error[Error Code:4012]
Turn off the AOC power, reconnect the cable, and turn on the AOC again. If the error appears even after checking the above, turn off the main power of the in... Show Detail
- No:5993
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/04 15:03
- Category: Error/Warning
-
Turn off the AOC power, reconnect the cable, and turn on the AOC again. If the error appears even after checking the above, turn off the main power of the in... Show Detail
- No:5992
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/04 15:01
- Category: Error/Warning
-
AOC set value out of range[Error Code:4010]
Turn off the AOC power, reconnect the cable, and turn on the AOC again. If the error appears even after checking the above, turn off the main power of the in... Show Detail
- No:5991
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/04 15:00
- Category: Error/Warning
-
AOC command is invalid[Error Code:4009]
Turn off the AOC power, reconnect the cable, and turn on the AOC again. If the error appears even after checking the above, turn off the main power of the in... Show Detail
- No:5990
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/04 14:59
- Category: Error/Warning
-
AOCm syringe counter warning [Error Code:9059 9060]
Replace the syringe and then reset the counter. <GC instrument monitor setting> Press HOME button > [Function] > [Diagnosis] >... Show Detail
- No:5948
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/07/08 14:02
- Category: Error/Warning