FAQ - Frequently Asked Questions
- Products Search > GC > GC-2030
Keyword Search
GC-2030
Frequently Referenced FAQ
FAQ within [ GC-2030 ]
-
Gas Selector# main control irregularity[Error Code:2358 2...
Check that the set pressure for the gas selector is lower than the supply pressure. Check that the supply pressure to the gas selector is within the r... Show Detail
- No:6004
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/07 17:01
- Category: Error/Warning
-
Gas Selector# purge control irregularity[Error Code:2360 ...
Check that the supply pressure to the gas selector is within the range below. Check that the gas supply pressure is stable. If the error appears repe... Show Detail
- No:6005
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 10:56
- Category: Error/Warning
-
Gas Selector# sub control irregularity[Error Code:2362 2363]
Check that the set pressure for the gas selector is lower than the supply pressure. Check that the supply pressure to the gas selector is within the range... Show Detail
- No:6006
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 10:58
- Category: Error/Warning
-
CARx purge valve not completely closed[Error Code:2048 20...
Check that the purge line outlet is not blinded. If the error appears repeatedly even after checking the above, turn off the main power of the instrument, st... Show Detail
- No:6007
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 11:00
- Category: Error/Warning
-
DetAPCn makeup valve not completely closed[Error Code:205...
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative. Show Detail
- No:6008
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 11:03
- Category: Error/Warning
-
DAMCn makeup set point not reached[Error Code:2272 2273 2...
Check that the gas main valve is open. Check that the gas supply pressure is within the range below. Check that the supply pressure is sufficient for the... Show Detail
- No:6009
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 11:06
- Category: Error/Warning
-
DAMCn reference set point not reached[Error Code:2276 227...
Check that the gas main valve is open. Check that the gas supply pressure is within the range below. Check that the supply pressure is sufficient for the... Show Detail
- No:6010
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 11:11
- Category: Error/Warning
-
DetAPCn makeup was overpressure[Error Code:2222 2225 2228...
Do not open or close the main valve of the supply gas during APC operation. Restart the GC with the gas supplied. Check that the makeup gas supply pressu... Show Detail
- No:6011
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 11:20
- Category: Error/Warning
-
DetAPCn hydrogen was overpressure[Error Code:2223 2226 22...
Do not open or close the main valve of the supply gas during APC operation. Restart the GC with the gas supplied. Check that the hydrogen gas supply pres... Show Detail
- No:6012
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 11:24
- Category: Error/Warning
-
DetAPCn air was overpressure[Error Code:2224 2227 2230 2233]
Do not open or close the main valve of the supply gas during APC operation. Restart the GC with the gas supplied. Check that the air gas supply pressure ... Show Detail
- No:6013
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 11:32
- Category: Error/Warning
111-120 of Total 225