FAQ - Frequently Asked Questions
- Products Search > GC > GC-2030
Keyword Search
GC-2030
Frequently Referenced FAQ
FAQ within [ GC-2030 ]
-
DetAPCn makeup valve not completely closed[Error Code:205...
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative. Show Detail
- No:6008
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 11:03
- Category: Error/Warning
-
CARx purge valve not completely closed[Error Code:2048 20...
Check that the purge line outlet is not blinded. If the error appears repeatedly even after checking the above, turn off the main power of the instrument, st... Show Detail
- No:6007
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 11:00
- Category: Error/Warning
-
Gas Selector# sub control irregularity[Error Code:2362 2363]
Check that the set pressure for the gas selector is lower than the supply pressure. Check that the supply pressure to the gas selector is within the range... Show Detail
- No:6006
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 10:58
- Category: Error/Warning
-
Gas Selector# purge control irregularity[Error Code:2360 ...
Check that the supply pressure to the gas selector is within the range below. Check that the gas supply pressure is stable. If the error appears repe... Show Detail
- No:6005
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 10:56
- Category: Error/Warning
-
Gas Selector# main control irregularity[Error Code:2358 2...
Check that the set pressure for the gas selector is lower than the supply pressure. Check that the supply pressure to the gas selector is within the r... Show Detail
- No:6004
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/07 17:01
- Category: Error/Warning
-
PURn was overpressure[Error Code:2218 2219 2220 2221]
Check that the purge line is not blinded. Do not open or close the main valve of the supply gas during AFC operation. Restart the GC with the gas supplie... Show Detail
- No:6003
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/07 17:03
- Category: Error/Warning
-
CARn was overpressure[Error Code:2214 2215 2216 2217]
Do not open or close the main valve of the supply gas during AFC operation. Restart the GC with the gas supplied. Check that the gas supply pressure is w... Show Detail
- No:6002
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 10:40
- Category: Error/Warning
-
CARm AFC set point not reached[Error Code:2130 2131 2204 ...
Check that the gas main valve is open. Check that the set values of the column inlet pressure and total flow are within the AFC control range. Check that... Show Detail
- No:6001
- Date of Issue:2021/05/18 15:36
- Last Updated:2023/12/13 13:22
- Category: Error/Warning
-
Analysis waiting[Error Code:5112]
Set [Prep Run Wait Time] longer. The PREP RUN is activated when using the following functions. Splitless (split in standby) Carrier gas saver (saved... Show Detail
- No:6000
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 10:19
- Category: Error/Warning
-
CARm calc. ratio out of range[Error Code:5075 5076 5108 5...
Change the split ratio so that the following equation is satisfied. (Total flow) = (Column flow) + (Column flow * Split ratio) + (Purge flow) Show Detail
- No:5999
- Date of Issue:2021/05/18 15:36
- Last Updated:2023/12/13 13:23
- Category: Error/Warning
111-120 of Total 225