SCD ozonier voltage was exceeded upper limit [Error Code:...
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement. Show Detail
- No:5878
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/04 13:48
- Category: Error/Warning
-
SCD fan error [Error Code:6121]
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement. Show Detail
- No:5877
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/04 13:47
- Category: Error/Warning
-
TCD signal was out of range [Error Code:4201]
Perform zero adjustment again after the TCD baseline becomes stable. If it does not work, press (HOME)-[Detector] and change to the optimum current value ... Show Detail
- No:5876
- Date of Issue:2021/05/18 15:37
- Category: Error/Warning
-
DET#1 2 3 4 FTD adjustment failed [Error Code:4149 4150 4...
Check that connection of electrode connector is correct and collector fixing knurled screws are tightened. MaintenanceHelp_FTD-Collector-Replacement.pdf ... Show Detail
- No:5875
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/04 14:09
- Category: Error/Warning
-
DET#1 2 3 4 DTCD overvoltage detected in cell power [Erro...
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement. Show Detail
- No:5874
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/04 14:09
- Category: Error/Warning
-
DET#1 2 3 4 FPD overcurrent detected [Error Code:4125 412...
Ensure that strong light does not leak in photomultiplier current from the connection. If the error cannot be resolved after you cut off the light with blac... Show Detail
- No:5873
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/04 14:08
- Category: Error/Warning
-
APCa-b PCB error [Error Code:23 24 25 26 27 28]
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement. Show Detail
- No:5872
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/04 13:34
- Category: Error/Warning
-
DetAPCn PCB error [Error Code:13 14 15 16]
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement. Show Detail
- No:5871
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/04 13:33
- Category: Error/Warning
-
APCn hydrogen flowing over the specified value[Error Code...
Do not open or close the main valve of the supply gas during APC operation. Restart the GC with the gas supplied. Check that the hydrogen gas supply pres... Show Detail
- No:6020
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 12:09
- Category: Error/Warning
-
DetAPCn DCG control irregularity[Error Code:2152 2153 215...
Check that the BID detector gas supply pressure is within the range below. Check that the supply pressure is sufficient for the BID detector gas to flow. ... Show Detail
- No:6019
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/07/06 08:11
- Category: Error/Warning