FAQ - Frequently Asked Questions
- Products Search > GC
Keyword Search
Frequently Referenced FAQ
FAQ within [ GC ]
-
Detected hydrogen leak[Error Code:51]
Stop the hydrogen gas supply. Check that the column connections have no leak. Check that the column is not broken. Check the tubing connections when usi... Show Detail
- No:6021
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/03 12:17
- Category: Error/Warning
-
The dual AOC version does not match with each other[Error...
If the ROM version is not the same, correct operation is not guaranteed. Turn off the main power of the instrument, stop gas supply, and contact your Shimadz... Show Detail
- No:6022
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/03 12:19
- Category: Error/Warning
-
SCD nitrogen control irregularity[Error Code:6167]
Check that the nitrogen gas supply pressure is within the range below. Check that the supply pressure is sufficient for the nitrogen gas to flow. Check t... Show Detail
- No:6023
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/03 12:28
- Category: Error/Warning
-
SCD hydrogen control irregularity[Error Code:6163]
Check that the hydrogen gas supply pressure is within the range below. Check that the supply pressure is sufficient for the hydrogen gas to flow. Check t... Show Detail
- No:6024
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/03 12:24
- Category: Error/Warning
-
SCD ozonizer oxygen flow is low[Error Code:6156]
Set flow of oxygen for ozone at 16.0 mL/min or higher. Show Detail
- No:6025
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/08/02 15:52
- Category: Error/Warning
-
SCD furnace temperature is low[Error Code:6155]
Set the furnace temperature at 750 °C or higher. Show Detail
- No:6026
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/03 12:30
- Category: Error/Warning
-
SCD PMT light was leaked[Error Code:6154]
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement. Show Detail
- No:6027
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/07 16:59
- Category: Error/Warning
-
SCD flow controler#2 was not connected[Error Code:6152]
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement. Show Detail
- No:6028
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/03 12:34
- Category: Error/Warning
-
SCD flow controler#1 was not connected[Error Code:6151]
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement. Show Detail
- No:6029
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/03 12:36
- Category: Error/Warning
-
SCD heater relay has been fixed to ON[Error Code:6149]
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement. Show Detail
- No:6030
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/03 12:37
- Category: Error/Warning
1-10 of Total 255