FAQ - Frequently Asked Questions
- Products Search > GC
Keyword Search
Frequently Referenced FAQ
FAQ within [ GC ]
-
xxx temperature exceeds the limit [Error Code:1002 1003 1...
Please follow the steps below; Press [Ignore] to close the error dialog. Change the maximum temperature value. Screen_Max-Temperature.pdf ... Show Detail
- No:5922
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/12 01:29
- Category: Error/Warning
-
CARm is not controlled [Error Code:2141 2142 2143 2213]
Turn on the carrier gas control or remove unused carrier gas line from the analytical line. How to set Flow control [On/OFF] Press (HOME) - [Injection ... Show Detail
- No:5923
- Date of Issue:2021/05/18 15:36
- Last Updated:2023/12/13 13:23
- Category: Error/Warning
-
xxx sensor use time warning [Error Code:9012 9013 9014 90...
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement. Show Detail
- No:5924
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/12 01:49
- Category: Error/Warning
-
Ignition finished(retried) [Error Code:9011]
Check the following items. ・Check for gas leaks. ・Verify the gas flows. ・Diagnoseignition pulse and ignition sequence by standard ... Show Detail
- No:5925
- Date of Issue:2021/05/18 15:36
- Category: Error/Warning
-
DET#n ignition failed. [Error Code:4203 4204 4205 4206]
Inspect the FID jet according to the following manual. MaintenanceHelp_FID-Collector-Inspection-n-Jet-Replacement.pdf Press(HOME) - [Detector] button on... Show Detail
- No:5926
- Date of Issue:2021/05/18 15:36
- Category: Error/Warning
- Brached FAQ
-
CARx insert counter warning [Error Code:9007 9008 9009 9053]
Replace the insert and then reset the counter. Insert replacement MaintenanceHelp_SPL-WBI-Glass-Insert-Replacement.pdf Counter reset &l... Show Detail
- No:5927
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/07/08 14:10
- Category: Error/Warning
-
CARx septum counter warning [Error Code:9004 9005 9006 9052]
Replace the septum and then reset the counter. Septum replacement MaintenanceHelp_Inj-Septum-Replacement.pdf Counter reset <GC i... Show Detail
- No:5928
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/07/08 14:08
- Category: Error/Warning
-
Fan use time warning [Error Code:9002]
Risk of current leakage, heat, smoke, and ignition increases. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sa... Show Detail
- No:5929
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/08 10:57
- Category: Error/Warning
-
COL CRG use time warning [Error Code:9000]
Replace the coolant and then reset the counter. <GC instrument monitor setting> Press HOME button > [Function] > [Diagnosis] >... Show Detail
- No:5930
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/08 11:00
- Category: Error/Warning
-
Clean up program start error [Error Code:4305]
Set the initial temperature of the column oven at higher values than room temperature by more than 2℃. (Example, Set the temperature at more than 27 ℃ whe... Show Detail
- No:5931
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/08 11:17
- Category: Error/Warning
71-80 of Total 255