FAQ - Frequently Asked Questions
- Products Search > GC
Keyword Search
Frequently Referenced FAQ
FAQ within [ GC ]
-
AOC-20s sample vial removed[Error Code:6021]
Check and perform the following. • Redo the teaching using the vial to be used for actual analysis. [In case of AOC-30i] AOC-30i_AOC-20sU_SamplerTea... Show Detail
- No:6065
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/16 09:24
- Category: Error/Warning
-
AOC-20s holding error[Error Code:6020]
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative. Show Detail
- No:6064
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/11 15:49
- Category: Error/Warning
-
AOC-20s vial returning error[Error Code:6019]
Check and perform the following. • Redo the teaching using the vial to be used for actual analysis. [In case of AOC-30i] AOC-30i_AOC-20sU_SamplerTea... Show Detail
- No:6063
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/16 09:22
- Category: Error/Warning
-
AOC-20s vial setting error[Error Code:6018]
Check and perform the following. • Check for any vial left from the previous sampling. • Redo the teaching using the vial to be used for actual analysis.... Show Detail
- No:6062
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/16 09:20
- Category: Error/Warning
-
AOC-20s can not start[Error Code:6017]
Press the auto injector [RESET] button. Show Detail
- No:6061
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/11 15:28
- Category: Error/Warning
-
AOC-20s up/down error[Error Code:6016]
Check and perform the following. • Check whether the initialization process at power-on can be completed. • Redo the teaching using the vial to be used for... Show Detail
- No:6060
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/16 09:20
- Category: Error/Warning
-
AOC-20s expansion and control error[Error Code:6015]
Redo the teaching using the vial to be used for actual analysis. [In case of AOC-30i] AOC-30i_AOC-20sU_SamplerTeaching.pdf [In case of AOC-20i] ... Show Detail
- No:6059
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/16 09:19
- Category: Error/Warning
-
AOC-20s rotating error[Error Code:6014]
Redo the teaching using thea vial to be used for actual analysis. [In case of AOC-30i] AOC-30i_AOC-20sU_SamplerTeaching.pdf [In case of AOC-20i] ... Show Detail
- No:6058
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/16 09:17
- Category: Error/Warning
-
AOCm vial sensor error[Error Code:6040 6045]
Check and perform the following. • Check whether the operating environment is appropriate. • Remove any dust or particles from the vial sensor. Show Detail
- No:6057
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/11 15:11
- Category: Error/Warning
-
AOCm turret step out[Error Code:6039 6044]
Turn off the power, turn the turret gently with your hands, and check the following: • No cable is caught in the turret's notch. • No part is interfering w... Show Detail
- No:6056
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/11 14:53
- Category: Error/Warning
211-220 of Total 255