FAQ - Frequently Asked Questions
- Products Search > GC
Keyword Search
Frequently Referenced FAQ
FAQ within [ GC ]
-
SCD WDT error[Error Code:6148]
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement. Show Detail
- No:6031
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/03 12:37
- Category: Error/Warning
-
SCD hydrogen primary pressure is not enough[Error Code:6147]
Check that the hydrogen gas supply pressure is within the range below. Check that the supply pressure is sufficient for the hydrogen gas to flow. Check t... Show Detail
- No:6032
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/03 12:41
- Category: Error/Warning
-
SCD oxygen primary pressure is not enough[Error Code:6146]
Check that the gas main valve is open. Check that the gas supply pressure is within the range below. Check that the supply pressure is sufficient for the... Show Detail
- No:6033
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/03 12:52
- Category: Error/Warning
-
DCG flow rate is not enough[Error Code:4213]
Check that the gas main valve is open. Check that the gas supply pressure is within the range below. Check that the supply pressure is sufficient for the... Show Detail
- No:6034
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/03 12:59
- Category: Error/Warning
-
DCG APC are not ready[Error Code:4212]
Check that the BID detector gas supply pressure is within the range below. Check that the supply pressure is sufficient for the BID detector gas to flow. ... Show Detail
- No:6035
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/10 09:36
- Category: Error/Warning
-
H2, AIR APC are not ready[Error Code:4207]
Check that the supply pressures of hydrogen gas and air are within the ranges below. Check that the supply pressures are sufficient for hydrogen gas and a... Show Detail
- No:6036
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/10 09:29
- Category: Error/Warning
-
Gas Selector# sub valve not completely closed[Error Code:...
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative. Show Detail
- No:6037
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/10 09:28
- Category: Error/Warning
-
Gas Selector# purge valve not completely closed[Error Cod...
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative. Show Detail
- No:6038
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/10 09:24
- Category: Error/Warning
-
Gas Selector# main valve not completely closed[Error Code...
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative. Show Detail
- No:6039
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/10 09:23
- Category: Error/Warning
-
DAMCn reference was overpressure[Error Code:2292 2293 229...
Check the supply gas pressure. If the supply pressure is correct, turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sale... Show Detail
- No:6040
- Date of Issue:2021/05/18 15:33
- Last Updated:2021/06/11 15:36
- Category: Error/Warning
11-20 of Total 253