FAQ - Frequently Asked Questions
- Products Search > GC
Keyword Search
Frequently Referenced FAQ
FAQ within [ GC ]
-
Processing, unable to start[Error Code:9054]
Wait for the analysis to finish before starting the next analysis. Show Detail
- No:5952
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/03 14:40
- Category: Error/Warning
-
SCD in pyrotube use time warn[Error Code:9073]
Replace the inner pyrotube and then reset the counter. ※Replace the inner pyrotube MaintenanceHelp_SCD-Inner-Pyrotube-Replacement.pdf ※ Rese... Show Detail
- No:5953
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/03 15:00
- Category: Error/Warning
-
SCD out pyrotube use time warn.[Error Code:9074]
Replace the outer pyrotube and then reset the counter. ※Replace the outer pyrotube MaintenanceHelp_SCD-Outer-Pyrotube-Replacement.pdf ※Reset... Show Detail
- No:5954
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/03 15:53
- Category: Error/Warning
-
SCD furnace abs. press warning[Error Code:9077]
Replace the inner pyrotube. ※Replace the inner pyrotube MaintenanceHelp_SCD-Inner-Pyrotube-Replacement.pdf ※Reset the counter [HOME] - [Det... Show Detail
- No:5955
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/03 15:50
- Category: Error/Warning
-
SCD exhaust for vacuum error[Error Code:6113]
Determine the cause of the leak by referring to the table below. If the error persists, turn off the main power of the instrument, stop gas supply, and co... Show Detail
- No:5956
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/03 17:16
- Category: Error/Warning
-
SCD unable to auto start[Error Code:9078]
If a leak check has failed,check the following and perform a leak checkagain. Screen_SCD-Leak-Check.pdf Cause Solution ... Show Detail
- No:5957
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/03 17:21
- Category: Error/Warning
-
Gas Selector# main was overpressure[Error Code:2364 2365]
Check the supply gas pressure. If the supply pressure is correct, turn off the main power of the instrument, stop gas supply, and contact your S... Show Detail
- No:5958
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/03 17:30
- Category: Error/Warning
-
Gas Selector# purge was overpressure[Error Code:2366 2367]
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative. Show Detail
- No:5959
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/04 09:48
- Category: Error/Warning
-
Gas Selector# sub was overpressure[Error Code:2368 2369]
Check the supply gas pressure. If the supply pressure is correct, turn off the main power of the instrument, stop gas supply, and contact your S... Show Detail
- No:5960
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/04 09:51
- Category: Error/Warning
-
Prep run wait time elapsed[Error Code:9062]
For manual injection analyses, introduce the sample within the [Prep RUN Wait Time]. For analyses with extended pre-run time, such as for HS warming, set [P... Show Detail
- No:5961
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/04 09:59
- Category: Error/Warning
181-190 of Total 255