SCD flow controler#1 error [Error Code:6177]
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative. Show Detail
- No:5936
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/09 14:28
- Category: Error/Warning
-
DET#1 2 3 4 failed to start plasma[Error Code:4208 4209 4...
• Check that the detector gas flow is appropriate. • Ignition may be difficult when gas replacement is insufficient immediately after tubing conne... Show Detail
- No:5935
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/09 14:17
- Category: Error/Warning
-
System is not ready [Error Code:9010]
Start the analysis after the system is ready. If this message appears when all the parameters including temperature and flow are ready, check the followin... Show Detail
- No:5934
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/08 12:50
- Category: Error/Warning
-
Pre-Run program time error [Error Code:4304]
Do not set thetotal time longer than 9999.99 min. The analysis can be performed even after this message is displayed, however, the program finishes afte... Show Detail
- No:5933
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/08 11:29
- Category: Error/Warning
-
Clean up program time error [Error Code:4303]
Do not set the total time longer than 9999.99 min. The analysis can be performed even after this message is displayed, however, the program finishes aft... Show Detail
- No:5932
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/08 11:22
- Category: Error/Warning
-
Clean up program start error [Error Code:4305]
Set the initial temperature of the column oven at higher values than room temperature by more than 2℃. (Example, Set the temperature at more than 27 ℃ whe... Show Detail
- No:5931
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/08 11:17
- Category: Error/Warning
-
COL CRG use time warning [Error Code:9000]
Replace the coolant and then reset the counter. <GC instrument monitor setting> Press HOME button > [Function] > [Diagnosis] >... Show Detail
- No:5930
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/08 11:00
- Category: Error/Warning
-
Fan use time warning [Error Code:9002]
Risk of current leakage, heat, smoke, and ignition increases. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sa... Show Detail
- No:5929
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/08 10:57
- Category: Error/Warning
-
CARx septum counter warning [Error Code:9004 9005 9006 9052]
Replace the septum and then reset the counter. Septum replacement MaintenanceHelp_Inj-Septum-Replacement.pdf Counter reset <GC i... Show Detail
- No:5928
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/07/08 14:08
- Category: Error/Warning
-
CARx insert counter warning [Error Code:9007 9008 9009 9053]
Replace the insert and then reset the counter. Insert replacement MaintenanceHelp_SPL-WBI-Glass-Insert-Replacement.pdf Counter reset &l... Show Detail
- No:5927
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/07/08 14:10
- Category: Error/Warning