FAQ - Frequently Asked Questions
- Products Search > GC
Keyword Search
Frequently Referenced FAQ
FAQ within [ GC ]
-
CARm AFC control irregularity [Error Code:2134 2135 2136 ...
Check that the supply pressure to the AFC is within the range below. Check that the supply pressure is sufficient for the AFC gas to flow. ... Show Detail
- No:5914
- Date of Issue:2021/05/18 15:37
- Last Updated:2023/12/13 13:24
- Category: Error/Warning
-
CARn hydrogen flowing over the specified value [Error Cod...
Ensure that the set values in programs such as Column Oven Temperature Program and Pressure Program are proper. When the set values in the programs are prope... Show Detail
- No:5915
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/12 01:19
- Category: Error/Warning
-
DetAPCn hydrogen flowing over the specified value [Error ...
Do not open or close the main valve of the supply gas during APC operation. Restart the GC with the gas supplied. Check that the hydrogen gas supply pres... Show Detail
- No:5916
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/12 01:21
- Category: Error/Warning
-
Flow# valve control irregularity [Error Code:2322 2323 23...
The valve of the flow controller or board may have failed. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/servic... Show Detail
- No:5917
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/12 01:24
- Category: Error/Warning
-
Flow# press/flow offset value is shifted [Error Code:2338...
An error may occur on the pressure sensor. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative. Show Detail
- No:5918
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/12 01:26
- Category: Error/Warning
-
Gas Selector# main set point not reached [Error Code:2346...
Check whether gas is supplied and whether gas is not leaking from connections. Gas leaks may be occurring in several locations (other than displayed site). Show Detail
- No:5919
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/12 01:27
- Category: Error/Warning
-
Gas Selector# purge set point not reached [Error Code:234...
Check whether gas is supplied and whether gas is not leaking from connections. Gas leaks may be occurring in several locations (other than displayed site). Show Detail
- No:5920
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/12 01:28
- Category: Error/Warning
-
SCD vacuum pump use time warn. [Error Code:9071]
Replace or maintain the vacuum pump and then reset the counter. SCD vacuum pump replacement MaintenanceHelp_SCD-Vacuum-Pump-Onspection.pdf Cou... Show Detail
- No:5949
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/09 16:40
- Category: Error/Warning
-
SCD ozone scrub. use time warn. [Error Code:9072]
The risk of ozone entering the exhaust from the vacuum pump increases. Replace the filling materials of the ozone scrubber and then reset the counter. ... Show Detail
- No:5950
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/09 16:51
- Category: Error/Warning
-
Not execute this operation [Error Code:9061]
The following operation was executed while theGC stopped. • Ignition/Illumination • FTD Feedback • Start analysis • Batch • Clean Up • PREP RUN • ... Show Detail
- No:5951
- Date of Issue:2021/05/18 15:37
- Category: Error/Warning
- Brached FAQ
171-180 of Total 255