APCx was overpressure[Error Code:2234 2235 2236 2237 2238...
Do not open or close the main valve of the supply gas during AUX-APC operation. Restart the GC with the gas supplied. Check that the gas supply pressure ... Show Detail
- No:6016
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 11:48
- Category: Error/Warning
-
DetAPCn ECDG was overpressure[Error Code:2268 2269 2270 2...
Do not open or close the main valve of the supply gas during APC operation. Restart the GC with the gas supplied. Check that the ECD gas supply pressure ... Show Detail
- No:6015
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 11:40
- Category: Error/Warning
-
DetAPCn DCG was overpressure[Error Code:2264 2265 2266 2267]
Do not open or close the main valve of the supply gas during APC operation. Restart the GC with the gas supplied. Check that the BID gas supply pressure ... Show Detail
- No:6014
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 11:34
- Category: Error/Warning
-
DetAPCn air was overpressure[Error Code:2224 2227 2230 2233]
Do not open or close the main valve of the supply gas during APC operation. Restart the GC with the gas supplied. Check that the air gas supply pressure ... Show Detail
- No:6013
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 11:32
- Category: Error/Warning
-
DetAPCn hydrogen was overpressure[Error Code:2223 2226 22...
Do not open or close the main valve of the supply gas during APC operation. Restart the GC with the gas supplied. Check that the hydrogen gas supply pres... Show Detail
- No:6012
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 11:24
- Category: Error/Warning
-
DetAPCn makeup was overpressure[Error Code:2222 2225 2228...
Do not open or close the main valve of the supply gas during APC operation. Restart the GC with the gas supplied. Check that the makeup gas supply pressu... Show Detail
- No:6011
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 11:20
- Category: Error/Warning
-
DAMCn reference set point not reached[Error Code:2276 227...
Check that the gas main valve is open. Check that the gas supply pressure is within the range below. Check that the supply pressure is sufficient for the... Show Detail
- No:6010
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 11:11
- Category: Error/Warning
-
DAMCn makeup set point not reached[Error Code:2272 2273 2...
Check that the gas main valve is open. Check that the gas supply pressure is within the range below. Check that the supply pressure is sufficient for the... Show Detail
- No:6009
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 11:06
- Category: Error/Warning
-
DetAPCn makeup valve not completely closed[Error Code:205...
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative. Show Detail
- No:6008
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 11:03
- Category: Error/Warning
-
CARx purge valve not completely closed[Error Code:2048 20...
Check that the purge line outlet is not blinded. If the error appears repeatedly even after checking the above, turn off the main power of the instrument, st... Show Detail
- No:6007
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 11:00
- Category: Error/Warning
131-140 of Total 253