DetAPCn DCG valve not completely closed[Error Code:2148 2...
Close the knob of the bypass unit of BID totally with the BID DCG flow set off. Then, check whether the BID DCG flow on the monitor decreases. When the BID D... Show Detail
- No:6018
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 11:56
- Category: Error/Warning
-
CARx prim is out of range[Error Code:2138 2139 2140 2212]
Check that the carrier gas supply pressure is within the primary pressure range in the configuration described below. Screen_Other-Configuration.pdf ... Show Detail
- No:6017
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 11:51
- Category: Error/Warning
-
APCx was overpressure[Error Code:2234 2235 2236 2237 2238...
Do not open or close the main valve of the supply gas during AUX-APC operation. Restart the GC with the gas supplied. Check that the gas supply pressure ... Show Detail
- No:6016
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 11:48
- Category: Error/Warning
-
DetAPCn ECDG was overpressure[Error Code:2268 2269 2270 2...
Do not open or close the main valve of the supply gas during APC operation. Restart the GC with the gas supplied. Check that the ECD gas supply pressure ... Show Detail
- No:6015
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 11:40
- Category: Error/Warning
-
DetAPCn DCG was overpressure[Error Code:2264 2265 2266 2267]
Do not open or close the main valve of the supply gas during APC operation. Restart the GC with the gas supplied. Check that the BID gas supply pressure ... Show Detail
- No:6014
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 11:34
- Category: Error/Warning
-
DetAPCn air was overpressure[Error Code:2224 2227 2230 2233]
Do not open or close the main valve of the supply gas during APC operation. Restart the GC with the gas supplied. Check that the air gas supply pressure ... Show Detail
- No:6013
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 11:32
- Category: Error/Warning
-
DetAPCn hydrogen was overpressure[Error Code:2223 2226 22...
Do not open or close the main valve of the supply gas during APC operation. Restart the GC with the gas supplied. Check that the hydrogen gas supply pres... Show Detail
- No:6012
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 11:24
- Category: Error/Warning
-
DetAPCn makeup was overpressure[Error Code:2222 2225 2228...
Do not open or close the main valve of the supply gas during APC operation. Restart the GC with the gas supplied. Check that the makeup gas supply pressu... Show Detail
- No:6011
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 11:20
- Category: Error/Warning
-
DAMCn reference set point not reached[Error Code:2276 227...
Check that the gas main valve is open. Check that the gas supply pressure is within the range below. Check that the supply pressure is sufficient for the... Show Detail
- No:6010
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 11:11
- Category: Error/Warning
-
DAMCn makeup set point not reached[Error Code:2272 2273 2...
Check that the gas main valve is open. Check that the gas supply pressure is within the range below. Check that the supply pressure is sufficient for the... Show Detail
- No:6009
- Date of Issue:2021/05/18 15:36
- Last Updated:2021/06/03 11:06
- Category: Error/Warning