Flow# DC24V was out of range [Error Code:59 62 65 68 71 74]
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement. Show Detail
- No:5888
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/12 00:29
- Category: Error/Warning
-
CARx calc. flow out of range [Error Code:5009 5010 5011 5...
Set the split ratio, column flow, and purge flow so that the total flow is within the AFC control range. Total flow is calculated with the following equat... Show Detail
- No:5887
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/12 00:25
- Category: Error/Warning
-
CARx calc. prss out of range [Error Code:5006 5007 5008 5...
Set the linear velocity and column flow so that the column inlet pressure is within the AFC control range. TechnicalDocument_AFCnAPC-Control-Ran... Show Detail
- No:5886
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/12 00:22
- Category: Error/Warning
-
Program time error [Error Code:4302]
Do not set the total time longer than 9999.99 min. Show Detail
- No:5885
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/12 00:13
- Category: Error/Warning
-
Settings were changed [Error Code:4301]
If you changed the part which is not executed, the instrument continues the analysis according to the changed program. Check the setting. Show Detail
- No:5884
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/04 13:58
- Category: Error/Warning
-
DET1 2 3 4 BID short-circuit in discharge cell [Error Cod...
Turn off the main power and then turn on the instrument again. If the error occurs repeatedly, cables may be damaged. Turn off the main power of the inst... Show Detail
- No:5883
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/04 13:56
- Category: Error/Warning
-
DAMCn makeup control irregularity [Error Code:2087 2088 2...
Check that the supply pressure to each unit is within the range below. Check that the gas supply pressure is stable. If the error appears rep... Show Detail
- No:5882
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/04 13:52
- Category: Error/Warning
-
xxx temperature controller error [Error Code:1037 1038 10...
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative. Show Detail
- No:5881
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/04 13:51
- Category: Error/Warning
-
Overheat is sensed [Error Code:1036]
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative. Show Detail
- No:5880
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/04 13:50
- Category: Error/Warning
-
SCD ozonier voltage fell below lower limit [Error Code:6123]
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement. Show Detail
- No:5879
- Date of Issue:2021/05/18 15:37
- Last Updated:2021/06/04 13:49
- Category: Error/Warning