1. Screw the compression load jig (3) into the load cell (2).
2. Loosen the set screw for the compression load jig temporarily and insert the bending punch (4).
3. Hold the bending punch (4) against the compression load jig (3) and tighten the set screw.
4. Fit the boss (7) of the bending support base (6) into the jig mounting surface (8) and fix with the hexagon socket head bolts (M8x2) (5).
NOTE:
• Screw in the compression load jig until the top surface of the load jig comes into contact with the bottom surface of the load cell.
• Tighten the set bolt (1) for the compression load jig with the top surface of the compression plate in contact with the compression load jig.