(AGX-V) The test force display is "- - - -"

(AGX-V) The test force display is "- - - -"
Category : 

Answer

• The load cell is not recognized. Turn off the power, check that the calibration cable is connected appropriately, and turn on the power again.
• If the type of display is "peak value" or "break point value", the value is displayed only after a test ends.