The operating status changes to [Error], and
is displayed.
Press [OK] to stop the alarm and return to the original screen.
• CONDUCTIVITY: A problem occurred in the detection unit.
• THERMISTOR: A problem occurred in the temperature sensor.
• ELECTRIC FAN ERROR: A problem occurred in the fan in the electric area.
• STORAGE: A problem occurred in the storage inside the instrument.
• INTERNAL SYSTEM: A problem occurred in the system.
• OPTION CONNECT: A problem occurred in the connection with the optional unit connected to the instrument.
→Turn off the power to the instrument and check whether the optional unit is correctly attached to the vial sampler terminal.
• OP1 COMMUNICATE: A problem occurred in the communication with the optional unit attached to vial sampler terminal (1).
→Turn off the power to the instrument and check whether the optional unit is correctly attached to the vial sampler terminal.
• OP2 COMMUNICATE: A problem occurred in the communication with the optional unit attached to vial sampler terminal (2).
→Turn off the power to the instrument and check whether the optional unit is correctly attached to the vial sampler terminal.
• INTERNAL MEMORY: A problem occurred in the data storage inside the instrument.