• No : 6822
  • Date of Issue : 2021/08/20 11:25
  • Print

(GAS) The indicator drifts (CGT-7100)

(GAS) The indicator drifts (CGT-7100)
Category : 

Answer

• There is a large change in ambient temperature. → Select a location with as little possible change in temperature.
• The inner surface of the cell is dirty. → Clean and purge the measurement cell. Contact a Shimadzu sales or service representative.
• Problem in sampling unit. → Perform checks and repairs on each part.
• Gas is leaking inside the analyzer. → Check the leak in the piping. Click here for "Gas leak checking (CGT-7100)".